Improvement of mechanical performance for vibratory microgyroscope based on sense mode closed-loop control

[+] Author Affiliations
Dingbang Xiao

National University of Defense Technology, College of Mechatronics and Automation, Changsha 410073, China

Jianbin Su, Zhihua Chen, Zhanqiang Hou, Xinghua Wang, Xuezhong Wu

National University of Defense Technology, College of Mechatronics and Automation, Changsha 410073, China

J. Micro/Nanolith. MEMS MOEMS. 12(2), 023001 (Apr 08, 2013). doi:10.1117/1.JMM.12.2.023001
History: Received August 25, 2012; Revised February 19, 2013; Accepted March 18, 2013
Text Size: A A A

Open Access Open Access

Abstract.  In order to improve its structural sensitivity, a vibratory microgyroscope is commonly sealed in high vacuum to increase the drive mode quality factor. The sense mode quality factor of the microgyroscope will also increase simultaneously after vacuum sealing, which will lead to a long decay time of free response and even self-oscillation of the sense mode. As a result, the mechanical performance of the microgyroscope will be seriously degraded. In order to solve this problem, a closed-loop control technique is presented to adjust and optimize the sense mode quality factor. A velocity feedback loop was designed to increase the electric damping of the sense mode vibration. A circuit was fabricated based on this technique, and experimental results indicate that the sense mode quality factor of the microgyroscope was adjusted from 8052 to 428. The decay time of the sense mode free response was shortened from 3 to 0.5 s, and the vibration-rejecting ability of the microgyroscope was improved obviously without sensitivity degradation.

Figures in this Article

The operation of a vibratory microgyroscope is based on thetransfer of energy from drive mode vibrations to sense mode vibrations caused by the Coriolis effect.1,2 The vibration amplitude of the drive mode mostly defines the structural sensitivity of the microgyroscope, so a higher drive mode quality factor is always desired to both improve rate precision and lower power consumption. The most effective technique to increase the drive mode quality factor is sealing the microgyroscope in high vacuum. However, in doing so, the sense mode quality factor will be increased simultaneously. A higher sense mode quality factor will lead to a longer decay time of the free response and even self-oscillation of the sense mode vibration. Consequently, external vibration will cause huge output errors, and the mechanical performance of the microgyroscope is seriously degraded.

A closed-loop control technique was presented to improve the mechanical performance of a vacuum-sealed microgyroscope. A velocity feedback loop was designed to increase the electric damping of sense mode vibrations. After analyzing the structural sensitivities of microgyroscopes with various sense mode quality factors, an optimized sense mode quality factor was suggested to improve the mechanical performance of the microgyroscope without sensitivity degradation. A feedback control circuit was fabricated, and experiments were carried out to validate this design.

The structural implementation of the microgyroscope designed in this paper is shown in Fig. 1(a).3 It includes two proof masses and a slanted suspension cantilever. The microstructure is fabricated by TMAH anisotropic etching. The slanted cantilever is formed by {100} and {111} crystal planes as shown in Fig. 1(b).4 The drive mode is the “see-saw” oscillation of the masses in the gyroscope plane, caused by the bending of the cantilever. The sense mode is the anti-phase oscillation of the masses out of the gyroscope plane, caused by the torsion of the cantilever. The drive and sense electrodes are fabricated on the glass substrate beneath the microstructure. The fabricated structure is shown in Fig. 1(c). The structure was sealed in a metal package at 1 Pa atmospheric pressure by cold welding as shown in Fig. 1(d).

Graphic Jump LocationF1 :

Structural diagram of the designed microgyroscope; (a) structural implementation; (b) cross-section of the cantilever; (c) fabricated structure; (d) sealed microgyroscope.

After vacuum sealing, the drive mode quality factor of this microgyroscope increased from 66 to 16,830, and the sense mode quality factor increased from 3 to 8052 simultaneously. The frequency responses of the fabricated microgyroscope were tested using frequency response analyzer FRA5087, and the results are shown in Fig. 2. The drive and sense mode frequencies were 2104 and 2053 Hz, respectively.

Graphic Jump LocationF2 :

The frequency responses of the vacuum-sealed microgyroscope.

The signal flow graph of this microgyroscope is shown in Fig. 3, where Dx and Dy are the DC voltages applied on the drive and sense electrodes, respectively, Axsinωxt is the sinusoidal driving voltage applied on the drive electrodes, Aysinωxt is the sinusoidal feedback voltage applied on the sense electrodes, Kvx and Kvy are voltage to moment conversion ratio of drive and sense axis, θp is the slanted angle of the cantilever of the microgyroscope, Ix and Iy are the inertia of the drive and sense mode, respectively, ωx and ωy are the resonant frequencies of drive and sense mode, respectively, Qx and Qy are the quality factors of drive and sense mode, respectively, Ωr is the input angular rate, φx and φy are the angular displacements of drive and sense mode vibration, respectively, Kcx and Kcy are displacement to capacitance conversion ratio of drive and sense axis, respectively, and ΔCx and ΔCy are the capacitive changes related to drive and sense mode vibration, respectively.

Graphic Jump LocationF3 :

Signal flow graph of the microgyroscope.

In keeping with a technique reported in many researches, a closed-loop control technique was used to generate drive mode resonance with constant amplitude. So the angular displacement of the drive mode vibration could be assumed as φx(t)=Pxcosωxt. If the step input angular rate is Ωr=AΩu(t), the differential equation for the sense mode angular displacement is Display Formula

φ¨y(t)+ωyQyφ˙y(t)+ωy2φy(t)=2AΩu(t)sinθpPxωxsinωxt.(1)

For the vacuum sealed microgyroscope, the solution to Eq. (1) is expressed as Display Formula

φy(t)=2AΩsinθpPxωxωy2(1ωx2ωy2)2+1Qy2ωx2ωy2sin(ωxtφl)2AΩsinθpPxωx2sinωyt(ωx2ωy2)ωyeωyt2Qy,(2)
where φl is the phase lag between the Coriolis force and sense mode vibration. φl is expressed as Display Formula
φl=tan1ωxωyQy(ωy2ωx2).(3)

The first term on the right-hand side of Eq. (2) is the forced response caused by the Coriolis effect. A coefficient kg is defined to express the structural static sensitivity of the microgyroscope as Display Formula

kg=1(1ωx2ωy2)2+(ωxωyQy)2.(4)

The relation ship between kg and Qy for various ωx/ωy is shown in Fig. 4. When the sense mode quality factor Qy is very small, the structural sensitivity coefficient kg increases quickly with an increase of Qy. Under the condition that Display Formula

Qy21(1ωx2ωy2)2ωx2ωy2(5)
the sensitivity coefficient kg will not increase with an increase of Qy anymore and tend to be constant.

Graphic Jump LocationF4 :

The dependence of the structural sensitivity kg on Qy.

The second term on the right-hand side of Eq. (2) is the free response of the sense mode. The time constant for the decay of the free response is Display Formula

τs=2Qyωy.(6)

A higher sense mode quality factor will lead to a longer decay time of the free response. For the fabricated microgyroscope in this paper, this time constant is 1.25s, which means that the output of the microgyroscope is very easily disturbed by external vibrations. In other words, the mechanical performance of the microgyroscope will degrade with the increase of the sense mode quality factor.

The sense mode quality factor should be optimized to improve the mechanical performance of the high vacuum-sealed microgyroscope. The quality factor decreases as the damping force increases. Sealing the microstructure at a higher atmospheric pressure is not a good choice because the drive mode quality factor will decrease as well as the sense mode quality factor. Designing a feedback control loop to increase the electric damping of the sense mode vibration seems to be the best choice. As the damping force is related to the velocity of structural movement, a velocity feedback control loop is required to generate an electric damping force on the vibrating structure.

The signal flow of the designed feedback control loop for the sense mode is shown in Fig. 5, where Fc is the Coriolis force, Kyv is the capacitance to voltage conversion ratio of sense axis, and Vyv is the output signal of the capacitance to voltage conversion circuit for the sense mode vibration. A differentiator is used to transfer the displacement signal to velocity signal and kfb is the feedback coefficient.

Graphic Jump LocationF5 :

Schematic diagram of the closed-loop control technique.

The transfer function from Fc to output voltage Vyv is Display Formula

Vyv(s)Fc(s)=KcyKyv/Iys2+ωyQys+ωy2(kvyDykfbKcvKyv/Iy)s.(7)

The modified sense mode quality factor of the feedback system shown in Fig. 5 is expressed as Display Formula

Qy=ωyωyQy(kvyDykfbKcyKyv/Iy).(8)

The sense mode quality factor should be decreased to improve the mechanical performance of the vacuum sealed microgyroscope, but the structural sensitivity of the microgyroscope will degrade dramatically when the sense mode quality factor is very low. The optimized sense mode quality factor Qsp is suggested to be Display Formula

Qsp=22|1ωx2ωy2|ωxωy.(9)

Under this condition, the sensitivity of microgyroscope only degrades 0.1% after using this feedback control technique. From Eqs. (8) and (9), the feedback coefficient kfb required for this optimized sense mode quality factor is determined by Display Formula

kfbop=Iy(ωyQyωy222ωx|1ωx2ωy2|)KvyDyKchKyv.(10)

The sense mode closed-loop controlling scheme is designed as shown in Fig. 6. The measurement of the sense capacitance is made by the two equal amplitude, out-of-phase sinusoidal modulation signals, +Vm and Vm, with the same frequency ωm. In the meantime, DC voltage Dy is added to the excitation signals in the sense electrodes. A differentiator is inserted between the low-pass filter and the amplifier Gfb. The output signals of the amplifier Gfb and the phase inverter following the amplifier are added to the top and bottom plates, respectively. An electric damping force will be generated on the sense axis, and the sense mode quality factor will be decreased.

Graphic Jump LocationF6 :

Sense mode closed-loop controlling technique.

The frequency responses of this closed-loop system with various gains Gfb were tested using FRA5087. The results are shown in Fig. 7(a). The relationship between quality factor and Gfb is shown in Fig. 7(b). It can be seen that the quality factor of the closed-loop system decreased with an increase of the velocity feedback coefficient.

Graphic Jump LocationF7 :

Frequency responses and quality factor of sense mode with closed-loop control; (a) frequency responses with various Gfb; (b) relationship between sense mode quality factor and Gfb.

According to Eq. (9), the optimized sense mode quality factor was suggested to be 449. Using the feedback coefficient determined by Eq. (10), the sense mode quality factor of the microgyroscope with closed-loop control was tested to be 428 as shown in Fig. 8, which was very close to the target value.

Graphic Jump LocationF8 :

Frequency response of the sense mode with closed-loop control.

When a shock is applied on the microgyroscope, the output signals of the capacitance to voltage conversion circuit for the sense mode vibration are shown in Fig. 9. The decay time required for the microgyroscope was shortened from 3 to 0.5 s after using this closed-loop control technique.

Graphic Jump LocationF9 :

The output signals of the capacitance to voltage conversion circuit for the sense mode vibration when applying a shock.

The outputs of this microgyroscope were measured under external random vibration. The frequency range of the random vibration was 20 to 2000Hz and the vibration intensity was 6grms. The output errors of the microgyroscope without closed-loop control increased dramatically during external vibrations, as shown in Fig. 10(a), which means that this open-loop microgyroscope can hardly work in a vibrating environment. After using this closed-loop control technique, the output noise of the microgyroscope during external vibration was almost unchanged, as shown in Fig. 10(b). It can be concluded that the vibration-rejecting ability of the microgyroscope was obviously improved by this closed-loop control technique.

Graphic Jump LocationF10 :

Outputs of this microgyroscope under external random vibration.

In order to improve the mechanical performance of a microgyroscope sealed in high vacuum, a closed-loop control technique based on velocity feedback was designed to modify the sense mode quality factor of the microgyroscope without structural sensitivity degradation. Experimental results indicated that the sense mode quality factor Qy was optimized from 8052 to 428, and the decay time of the sense mode free response was shortened from 3 to 0.5 s. The vibration-rejecting ability of the microgyroscope was obviously improved by this closed-loop control technique.

This work was supported by National Natural Science Foundation of China (Grant No. 51005239) and National University of Defense Technology Foundation (No. JC12-03-04). The authors would like to thank the Laboratory of Microsystem, National University of Defense Technology, China, for equipment access and technical support.

Trusov  A. A., Schofield  A. R., Shkel  A. M., “Performance characterization of a new temperature-robust gain-bandwidth improved MEMS gyroscope operated in air,” Sens. Actuat. A. 155, (1 ), 16 –22 (2009). 0924-4247 CrossRef
Tsai  N.-C., Sue  C.-Y., Lin  C.-C., “Micro angular rate sensor design and nonlinear dynamics,” J. Micro/Nanolith. MEMS MOEMS. 6, (3 ), 033008  (2007). 1932-5134 CrossRef
Dingbang  X., “High-performance micromachined gyroscope with a slanted suspension cantilever,” J. Semiconduct.. 30, (4 ), 044012  (2009). 1674-4926 CrossRef
Hou  Z., “Effect of axial force on the performance of micromachined vibratory rate gyroscopes,” Sensors. 11, (1 ), 296 –309 (2011). 0746-9462 CrossRef

Grahic Jump LocationImage not available.

Dingbang Xiao is an associate professor. He received his PhD in mechanical engineering from the National University of Defense Technology, China, in 2009. His research interests include mechatronics engineering, MEMS, and weak signal process.

Biographies and photographs of the other authors are not available.

© The Authors. Published by SPIE under a Creative Commons Attribution 3.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.

Citation

Dingbang Xiao ; Jianbin Su ; Zhihua Chen ; Zhanqiang Hou ; Xinghua Wang, et al.
"Improvement of mechanical performance for vibratory microgyroscope based on sense mode closed-loop control", J. Micro/Nanolith. MEMS MOEMS. 12(2), 023001 (Apr 08, 2013). ; http://dx.doi.org/10.1117/1.JMM.12.2.023001


Figures

Graphic Jump LocationF1 :

Structural diagram of the designed microgyroscope; (a) structural implementation; (b) cross-section of the cantilever; (c) fabricated structure; (d) sealed microgyroscope.

Graphic Jump LocationF2 :

The frequency responses of the vacuum-sealed microgyroscope.

Graphic Jump LocationF3 :

Signal flow graph of the microgyroscope.

Graphic Jump LocationF4 :

The dependence of the structural sensitivity kg on Qy.

Graphic Jump LocationF5 :

Schematic diagram of the closed-loop control technique.

Graphic Jump LocationF6 :

Sense mode closed-loop controlling technique.

Graphic Jump LocationF7 :

Frequency responses and quality factor of sense mode with closed-loop control; (a) frequency responses with various Gfb; (b) relationship between sense mode quality factor and Gfb.

Graphic Jump LocationF8 :

Frequency response of the sense mode with closed-loop control.

Graphic Jump LocationF9 :

The output signals of the capacitance to voltage conversion circuit for the sense mode vibration when applying a shock.

Graphic Jump LocationF10 :

Outputs of this microgyroscope under external random vibration.

Tables

References

Trusov  A. A., Schofield  A. R., Shkel  A. M., “Performance characterization of a new temperature-robust gain-bandwidth improved MEMS gyroscope operated in air,” Sens. Actuat. A. 155, (1 ), 16 –22 (2009). 0924-4247 CrossRef
Tsai  N.-C., Sue  C.-Y., Lin  C.-C., “Micro angular rate sensor design and nonlinear dynamics,” J. Micro/Nanolith. MEMS MOEMS. 6, (3 ), 033008  (2007). 1932-5134 CrossRef
Dingbang  X., “High-performance micromachined gyroscope with a slanted suspension cantilever,” J. Semiconduct.. 30, (4 ), 044012  (2009). 1674-4926 CrossRef
Hou  Z., “Effect of axial force on the performance of micromachined vibratory rate gyroscopes,” Sensors. 11, (1 ), 296 –309 (2011). 0746-9462 CrossRef

Some tools below are only available to our subscribers or users with an online account.

Related Content

Customize your page view by dragging & repositioning the boxes below.

Related Journal Articles

Related Book Chapters

Topic Collections

Advertisement
  • Don't have an account?
  • Subscribe to the SPIE Digital Library
  • Create a FREE account to sign up for Digital Library content alerts and gain access to institutional subscriptions remotely.
Access This Article
Sign in or Create a personal account to Buy this article ($20 for members, $25 for non-members).
Access This Proceeding
Sign in or Create a personal account to Buy this article ($15 for members, $18 for non-members).
Access This Chapter

Access to SPIE eBooks is limited to subscribing institutions and is not available as part of a personal subscription. Print or electronic versions of individual SPIE books may be purchased via SPIE.org.