Paper
8 October 1992 X-ray imaging with the XSPECT/SODART microstrip proportional counters
Carl Budtz-Joergensen, Axel Bahnsen, M. Mohl Madsen, C. Olesen, Herbert W. Schnopper
Author Affiliations +
Abstract
The Danish Space Research Institute will provide a set of four imaging microstrip proportional counters (MSPC) as part of XSPECT, the Danish contribution to the SODART telescopes. A high- and a low-energy detector (HEPC and LEPC) will be provided for each of the SODART telescopes. The design is fixed and a prototype of LEPC is under construction. The present paper describes the imaging properties of this detector. The inherent position resolution is < 0.2 mm around 6 keV, and positional nonlinearities are < 0.1 mm. The gas gain is independent of position to better than 0.5%. Finally, the point spread function, important for studying weak objects in the vicinity of strong ones, has been measured.
© (1992) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Carl Budtz-Joergensen, Axel Bahnsen, M. Mohl Madsen, C. Olesen, and Herbert W. Schnopper "X-ray imaging with the XSPECT/SODART microstrip proportional counters", Proc. SPIE 1743, EUV, X-Ray, and Gamma-Ray Instrumentation for Astronomy III, (8 October 1992); https://doi.org/10.1117/12.130678
Lens.org Logo
CITATIONS
Cited by 6 scholarly publications.