Abstract
In Fourier transform infrared spectrometer (FTIR), the optical path difference of laser metrology is applied to increase spectrum measure precision. However, due to the characteristics of laser single frequency and steady frequency, the uneven speed of moving mirror, circuit delay etc. cause the deviation of sampling point. In this paper, based on the interference theory of spectrometer, according to the Fourier contrary transform and methods of error analysis, the theory model between relative error of spectrum measure and the deviation of sampling point is established. Some simulation computations of the model have been done, the result indicates that, as a theory basis, the model can be applied in analyzing the sampling error of spectrometer and correction algorithm.
© (2013) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Ya Kang "Analysis of sampling error in FTIR", Proc. SPIE 8761, PIAGENG 2013: Image Processing and Photonics for Agricultural Engineering, 87610N (4 March 2013); https://doi.org/10.1117/12.2019626