9 March 2017 A virtual clinical trial using projection-based nodule insertion to determine radiologist reader performance in lung cancer screening CT
Lifeng Yu, Qiyuan Hu, Chi Wan Koo, Edwin A. Takahashi, David L. Levin, Tucker F. Johnson, Megan J. Hora, Shane Dirks, Baiyu Chen, Kyle McMillan, Shuai Leng, J. G. Fletcher, Cynthia H. McCollough
Author Affiliations +
Abstract
Task-based image quality assessment using model observers is promising to provide an efficient, quantitative, and objective approach to CT dose optimization. Before this approach can be reliably used in practice, its correlation with radiologist performance for the same clinical task needs to be established. Determining human observer performance for a well-defined clinical task, however, has always been a challenge due to the tremendous amount of efforts needed to collect a large number of positive cases. To overcome this challenge, we developed an accurate projection-based insertion technique. In this study, we present a virtual clinical trial using this tool and a low-dose simulation tool to determine radiologist performance on lung-nodule detection as a function of radiation dose, nodule type, nodule size, and reconstruction methods. The lesion insertion and low-dose simulation tools together were demonstrated to provide flexibility to generate realistically-appearing clinical cases under well-defined conditions. The reader performance data obtained in this virtual clinical trial can be used as the basis to develop model observers for lung nodule detection, as well as for dose and protocol optimization in lung cancer screening CT.
© (2017) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Lifeng Yu, Qiyuan Hu, Chi Wan Koo, Edwin A. Takahashi, David L. Levin, Tucker F. Johnson, Megan J. Hora, Shane Dirks, Baiyu Chen, Kyle McMillan, Shuai Leng, J. G. Fletcher, and Cynthia H. McCollough "A virtual clinical trial using projection-based nodule insertion to determine radiologist reader performance in lung cancer screening CT", Proc. SPIE 10132, Medical Imaging 2017: Physics of Medical Imaging, 101321R (9 March 2017); https://doi.org/10.1117/12.2255593
Lens.org Logo
CITATIONS
Cited by 4 scholarly publications and 1 patent.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Computed tomography

Lung

Lung cancer

Performance modeling

Image quality

Lung imaging

Back to Top