A low noise superheterodyne receiver has been designed and constructed for the ground-based monitoring of stratospheric ozone. The receiver measures the ozone spectral line at 110.836 GHz. The low-noise front-end consists of a Schottky-mixer, a HEMT-amplifier at 3.6 GHz and a phase-locked Gunn-oscillator as LO and is followed by an acousto-optical dual spectrometer with 1.2 GHz bandwidth and 85 kHz resolution at the band center. A tunable passive cold load is used as a reference load.
© (1990) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Heikki Valmu "A 110.8 GHz radiometer for stratospheric ozone monitoring", Proc. SPIE 1514, 15th International Conference on Infrared and Millimeter Waves, 151434 (1 December 1990); https://doi.org/10.1117/12.2301522