30 May 1997 Aerosol fluorescence spectrum analyzer for individual airborne biological particles
Gang Chen, Ronald G. Pinnick, Steven C. Hill, Richard K. Chang, Paul Nachman, Gorden W. Videen
Author Affiliations +
Abstract
We report the operation of an aerosol analyzer capable of measuring the fluorescence spectra of single micrometer-size bioaerosol particles. Aerosol particles in an air stream initially transverse a cw 488-nm 'trigger' laser beam where their elastic scattering and total fluorescence is measured with photomultipliers. When the elastic scattering and fluorescence signals meet certain criteria, a UV (266 nm) 'probe' laser is triggered and it illuminates the selected particles. The UV laser-excited spectra of particles are measured with the instrument's image-intensified CCD detector, gaged with signals from the trigger laser. We demonstrate the ability of the instrument to capture the fluorescence of single airborne biological particles. The results suggest that it may be possible to differentiate among biological particles based on their single particle fluorescence spectra.
© (1997) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Gang Chen, Ronald G. Pinnick, Steven C. Hill, Richard K. Chang, Paul Nachman, and Gorden W. Videen "Aerosol fluorescence spectrum analyzer for individual airborne biological particles", Proc. SPIE 3105, Chemical, Biochemical and Environmental Fiber Sensors IX, (30 May 1997); https://doi.org/10.1117/12.276175
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Luminescence

Atmospheric particles

Particles

Aerosols

Laser scattering

Laser beam diagnostics

Scattering

Back to Top