Paper
6 October 2003 Measurement of signal-to-noise ratio for mixture of coherent and incoherent sources using digital sampling oscilloscope
Gyeon-il Kweon, Jung-ho Choi, Ji-heon Jeong
Author Affiliations +
Proceedings Volume 5137, International Conference on Lasers, Applications, and Technologies 2002: Advanced Lasers and Systems; (2003) https://doi.org/10.1117/12.517935
Event: International Conference on Lasers, Applications, and Technologies 2002, 2002, Moscow, Russian Federation
Abstract
Signal-to-noise for a coherent laser source and a mixture of a laser source with a narrow band amplified spontaneous emission source are directly measured using a digital sampling oscilloscope. The measurement for a coherent source is found unreliable due to the low level of shot-noise compared to the thermal noise. On the other hand, the measurement for a mixed source is found in good agreement with theoretical predictions within 1dB error.
© (2003) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Gyeon-il Kweon, Jung-ho Choi, and Ji-heon Jeong "Measurement of signal-to-noise ratio for mixture of coherent and incoherent sources using digital sampling oscilloscope", Proc. SPIE 5137, International Conference on Lasers, Applications, and Technologies 2002: Advanced Lasers and Systems, (6 October 2003); https://doi.org/10.1117/12.517935