2 March 2006 Calculation of images from an anthropomorphic chest phantom using Monte Carlo methods
Gustaf Ullman, Alexandr Malusek, Michael Sandborg, David R. Dance, Gudrun Alm Carlsson
Author Affiliations +
Abstract
Monte Carlo (MC) computer simulation of chest x-ray imaging systems has hitherto been performed using anthropomorphic phantoms with too large (3 mm) voxel sizes. The aim for this work was to develop and use a Monte Carlo computer program to compute projection x-ray images of a high-resolution anthropomorphic voxel phantom for visual clinical image quality evaluation and dose-optimization. An Alderson anthropomorphic chest phantom was imaged in a CT-scanner and reconstructed with isotropic voxels of 0.7 mm. The phantom was segmented and included in a Monte Carlo computer program using the collision density estimator to derive the energies imparted to the detector per unit area of each pixel by scattered photons. The image due to primary photons was calculated analytically including a pre-calculated detector response function. Attenuation and scatter of x-rays in the phantom, grid and image detector was considered. Imaging conditions (tube voltage, anti-scatter device) were varied and the images compared to a real computed radiography (Fuji FCR 9501) image. Four imaging systems were simulated (two tube voltages 81 kV and 141 kV using either a grid with ratio 10 or a 30 cm air gap). The effect of scattered radiation on the visibility of thoracic vertebrae against the heart and lungs is demonstrated. The simplicity in changing the imaging conditions will allow us not only to produce images of existing imaging systems, but also of hypothetical, future imaging systems. We conclude that the calculated images of the high-resolution voxel phantom are suitable for human detection experiments of low-contrast lesions.
© (2006) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Gustaf Ullman, Alexandr Malusek, Michael Sandborg, David R. Dance, and Gudrun Alm Carlsson "Calculation of images from an anthropomorphic chest phantom using Monte Carlo methods", Proc. SPIE 6142, Medical Imaging 2006: Physics of Medical Imaging, 614239 (2 March 2006); https://doi.org/10.1117/12.644119
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Monte Carlo methods

Imaging systems

Sensors

Photons

Chest

Signal to noise ratio

Lung

Back to Top