Abstract
A method of reducing measuring time of temperature measurements of biological objects based on preheating the resistance temperature detector (RTD) up to the temperature close to the temperature to be measured, is proposed. It has been found that at the same measuring time, the preheating allows to decrease the measurement error by a factor of 5 to 45 over the temperature range of 35-41°С. The measurement time is reduced by 1.6-4 times over this range, keeping the same value of the measurement error.
© (2014) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
O. Hotra, O. Boyko, and T. Zyska "Improvement of the operation rate of medical temperature measuring devices", Proc. SPIE 9291, 13th International Scientific Conference on Optical Sensors and Electronic Sensors, 92910A (19 August 2014); https://doi.org/10.1117/12.2070167