1 July 2006 Quantitative evaluation of optical coherence tomography signal enhancement with gold nanoshells
Author Affiliations +
Abstract
Nanoshell-enhanced optical coherence tomography (OCT) is a novel technique with the potential for molecular imaging and improved disease detection. However, optimization of this approach will require a quantitative understanding of the influence of nanoshell parameters on detected OCT signals. In this study, OCT was performed at 1310 nm in water and turbid tissue-simulating phantoms to which nanoshells were added. The effect of nanoshell concentration, core diameter, and shell thickness on signal enhancement was characterized. Experimental results indicated trends that were consistent with predicted optical properties—a monotonic increase in signal intensity and attenuation with increasing shell and core size. Threshold concentrations for a 2-dB OCT signal intensity gain were determined for several nanoshell geometries. For the most highly backscattering nanoshells tested—291-nm core diameter, 25-nm shell thickness—a concentration of 109 nanoshells/mL was needed to produce this signal increase. Based on these results, we discuss various practical considerations for optimizing nanoshell-enhanced OCT. Quantitative experimental data presented here will facilitate optimization of OCT-based diagnostics and may also be relevant to other reflectance-based approaches as well.

1.

Introduction

Optical coherence tomography (OCT) has received considerable attention as a revolutionary biomedical imaging method since its introduction just over a decade ago.1 However, only recently have researchers begun to explore the possibilities of improving OCT image quality in biological tissue with exogenous contrast agents, many of which can provide molecular contrast. 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13 For example, air-filled microbubbles2 and engineered microspheres3 have been used to enhance the intensity of backscattered light from the tissue for OCT imaging. Near-infrared dyes that are spectrally active have been shown to enhance spectroscopic OCT images.4

Metal nanoparticles—including homogeneous nanospheres,14 nanoshells,15, 16, 17 and nanocages13, 18—represent a new generation of contrast agents that show promise for improving signals in reflectance-based diagnostics. Nanoshells consist of a dielectric (silica) core covered by a thin metallic shell. While multilayer nanoshells16 have been investigated theoretically, construction of such particles has not been achieved. The current study involves nanoshells with a single gold layer. By varying the relative dimensions of core and shell, the optical resonance of these particles can be systematically varied over a broad spectral region from the near-uv to the mid-ir. Therefore, it is possible to engineer gold nanoshells for OCT imaging that have a high level of backscattering and a low level of absorption at the appropriate wavelength. Another important property of nanoshells is their ability to be bioconjugated to enzymes and antibodies for selective targeting to receptors of clinical interest.14, 19 In the future, this line of research may lead to an OCT-based molecular imaging approach that is sensitive and specific to neoplastic lesions and provides high levels of contrast.

OCT imaging with gold nanoshells has been documented. Loo 15 compared in vitro OCT images of water, microsphere suspension, and nanoshell suspension at an illumination wavelength of 830nm . The nanoshells used in this study had a silica core diameter of 200-nm and 20-nm shell thickness and were suspended in water at a concentration of 109mL . Experimental data on the influence of specific nanoshell parameters are not yet available in the literature. Such data are needed to elucidate the optical characteristics of these particles and facilitate the development and optimization of nanoshell-enhanced OCT for minimally invasive diagnostics. Therefore, the goal of this study was to quantify the effect of nanoshell core diameter, shell thickness and concentration on the contrast enhancement provided by nanoshells during OCT imaging at 1310nm .

2.

Methods

Optical properties of nanoshells were determined from analytical vector-based solutions to Maxwell’s equations for the incident, internal, and scattered electromagnetic fields.20, 21, 22 Briefly, matching the boundary conditions between the core and shell and between the shell and surrounding medium allows coefficients defining the electric and magnetic fields to be determined. These coefficients are dependent on the dielectric constants of the core, shell, and medium, on the core and total nanoshell radii, and on the wavelength. Two of these coefficients, an and bn , define the scattered field and therefore permit direct computation of the extinction, scattering, and backscattering efficiencies ( Qext , Qsca , Qback ), as follows23:

Qext=2x2n=1(2n+1)Re{an+bn},
Qsca=2x2n=1(2n+1)(an2+bn2),
Qback=1x2n=1(2n+1)(1)n(anbn)2,
where x is the size parameter, 2πrλ ; r is the total nanoshell radius; and λ is the wavelength in the surrounding medium. Absorption efficiency (Qabs) is then determined from Qext and Qsca : Qabs=QextQsca . Each optical transport coefficient ( μx cm1 ) is easily computed from its corresponding efficiency (Qx) by μx=Qxπr2ρ , where ρ is the volume density of particles (cm3) . At an illumination wavelength of 1310nm , calculations indicated that nanoshells with a core diameter over 200nm and shell thickness greater than 15nm show higher scattering efficiencies. In addition, absorption efficiencies decrease dramatically as shell thickness increases over approximately 15nm . Thus, nanoshells with these size characteristics were expected to yield strong OCT signal.

Nanoshells were fabricated by the authors at Rice University. The fabrication protocol developed for nanoshells includes molecular self-assembly and colloid chemistry in an aqueous solution.24 For the nanoshell cores, silica nanoparticles were made by the Stöber method, which involves the reduction of tetraethylorthosilicate (Sigma-Aldrich, St. Louis, Missouri) in ethanol. The sizes of the silica cores were determined by scanning electron microscopy (SEM). The cores were functionalized with aminopropyltriethoxysilane (Sigma-Aldrich), resulting in attached amine groups. For the shell, colloidal gold nanoparticles of 1- to 3-nm diameter were fabricated according to the method of Duff 25 This colloidal suspension was first aged for two weeks at 4°C and concentrated using a rotary evaporator. When the functionalized silica cores were placed in the aged colloidal suspension, gold nanoparticles adsorbed to the amine groups. Adsorbed gold provided the nucleation sites for further reduction of gold, until these sites coalesced into a complete shell. The amount of gold added during this final reduction stage determined the shell thickness. Final gold nanoshell sizes were evaluated by SEM. Prior to OCT experiments, nanoshell surfaces were modified with polyethylene glycol (PEG)-thiol to provide steric stabilization as well as to eliminate nonspecific protein adsorption. PEG-thiol was synthesized by reacting 2-iminothiolane (Sigma-Aldrich) with PEG-NH2 5000 MW (Nektar Therapeutics, Huntsville, Alabama). After mixing and reacting equal volumes of the PEG-amine and iminothiolane for 1h , they were dialyzed against deionized water for 1 to 2h , changing the dialysate four times in order to remove excess reagent. While initial research indicates that PEGylation can decrease scattering intensity from nanoshells, we investigated only PEGylated nanoshells because these are the most relevant for in vivo applications. Nanoshells with six unique geometries were fabricated: (core diameter/shell thickness, in nanometers) 126/15, 126/26, 213/19, 291/8, 291/15, 291/25. Figure 1 shows a SEM image of a 291/15 nanoshell.

Fig. 1

SEM image of nanoshell with 291-nm core diameter and 15-nm shell thickness.

041121_1_037604jbo1.jpg

Nanoshell concentrations were determined via an approach resembling hemacytometer-based cell counting. Nanoshell suspensions were first diluted by a known factor to achieve 107 particles/mL. We then captured digital charge-coupled device (CCD) images of each diluted suspension in a 0.1-mm pathlength cuvette under a darkfield microscope with a 20× objective. Darkfield microscopy was preferred over brightfield because of substantially better image contrast. The CCD field of view was calibrated (microns/pixel) with a scale having 10μm rulings; therefore, the volume in the field of view was known. We used our own MATLAB code to locate and tally all the particles in each image and then compute the concentration. The precision of this counting approach, evaluated by counting three samples of each nanoshell suspension, was 12% or better (one standard deviation). The final estimates of nanoshell concentrations were accurate to within 9%, as determined by using the above method to count microsphere suspensions with known concentrations.

Measurements were performed with a time-domain OCT system based on an all-fiber common-path interferometer, also known as an autocorrelator (AIF-INST-02, Optiphase, Van Nuys, California).26 Sharma 27 have described and analyzed a similar OCT system configuration. A diagram of our system is shown in Fig. 2 . As compared to the standard Michelson interferometer, the autocorrelator does not have a separate reference arm and instead derives its reference signal from a reflection in the sample path. Depth ranging is achieved by two piezoelectric fiber stretchers driven by triangular voltage waveforms 180deg out of phase with each other. This push-pull operation permits depth scans up to 7mm in air. The OCT sample path consists of the single-mode fiber input/output of the autocorrelator coupled to an 11-mm focal length collimating lens (F220FC-C, Thorlabs, Newton, New Jersey), followed by x - and y -scanning galvanometer mirrors (Model 6210, Cambridge Technology, Cambridge, Massachusetts), and finally a refocusing lens (C280TM-C, Thorlabs) with an 18-mm focal length. Because the fiber has angled faces and the lenses have efficient antireflective coatings, the first possible back reflection that provides the reference signal occurs at the sample itself. To optimize signal-to-noise ratio (SNR), the magnitude of the reference can be adjusted by the type of cover glass (or lack thereof) over the sample and/or by the axial position of the beam focus. The shot noise-limited SNR of the system is 61dB . The system’s decibel intensity units were calibrated by recording the changes in signal intensity from a glass plate as the light source output power, measured with a U.S. National Institute of Standards and Technology–traceable power meter, was varied over two orders of magnitude. The light source for this OCT system is a superluminescent diode centered at 1310nm with a 50-nm spectral bandwidth, which therefore yields a coherence length of 17μm in air. The lateral resolution is 15μm at focus, and the optical power at the sample is approximately 12mW .

Fig. 2

Diagram of autocorrelator-based OCT system.

041121_1_037604jbo2.jpg

OCT measurements were taken of the nanoshells in water and turbid, tissue-simulating phantoms. The phantoms consisted of 1.53μm diameter polystyrene microspheres (Polysciences, Warrington, Pennsylvania) suspended in water. These phantoms had a scattering coefficient (μs) of 100cm1 and anisotropy of 0.9, based on Mie theory calculations. This level of scattering is directly relevant to biological tissue at 1310nm .28 Samples were imaged in a 0.5- or 1-mm pathlength two-piece cuvette, with the OCT beam axially focused at the interface between the cuvette glass and liquid sample, to maximize signal, and hence SNR, from the sample itself. Each B-scan image required 3s of acquisition time and consisted of 120 A-scans taken at 40Hz over a 0.85-mm lateral region. To ensure a homogeneous distribution of particles, each suspension was agitated just prior to OCT imaging. The particle settling time was analyzed and found to produce noticeable changes in the OCT signal about 45s after ceasing agitation. Therefore, each B-scan was acquired within 15s of agitation. Figures 3a and 3b show typical B-scan images of water and nanoshells in water. To reduce the effects of speckle and to obtain more reliable intensity estimates at low nanoshell concentrations, all 120 A-scans from each B-scan were averaged for subsequent analysis.

Fig. 3

OCT B-scan images of 1-mm pathlength cuvette filled with (a) water and (b) 291/25 nanoshells at a concentration of 5×109 particles/mL in water.

041121_1_037604jbo3.jpg

Two sets of measurements were performed in this study. The first set involved OCT imaging of nanoshells in water and phantom solutions where the concentration of nanoshells in each sample was held constant at 5×109 particles/mL. In the second set of measurements, we investigated the effect of nanoshell concentration on OCT signal intensity in water and phantom. Each data point displayed in the figures represents the mean of three measurements. The standard deviation for each data point was also computed, but the error bars on each graph represent the maximum standard deviation of all points on that graph.

3.

Results

The intent of the first set of measurements was to determine the intensity gain achieved by the addition of nanoshells with differing geometries. Intensity gain was defined as the average intensity over the first 100μm of a sample (water or phantom) with nanoshells minus the corresponding average intensity of that sample without nanoshells. The averaging was limited to the first 100μm beyond the glass-liquid interface (75 to 175μm from the peak) to minimize the effect of signal attenuation. For the case of nanoshells added to water, a graph of average A-scans for two nanoshell types and a water baseline is shown in Fig. 4a . Figure 4b shows the intensity gain values for all the nanoshells added to water. This graph shows a variation of intensity gains from 19 to 28dB in a monotonic relationship with the predicted backscattering coefficient (μback) . As expected, larger core and shell sizes tended to generate higher levels of backscattering and thus greater intensity gain. Note that the nanoshells with the largest core but smallest shell (291/8) produced a smaller signal than those with smaller core and larger shell (213/19).

Fig. 4

(a) Average A-scans of water and two different sizes of nanoshells in water ( 5×109 nanoshells/mL). The vertical dotted lines indicate the first 100μm over which the intensity was averaged for intensity gain estimates. (b) Measured intensity gain from each size of nanoshells versus the backscattering coefficient (μback) predicted for each size of nanoshells. Note that data points are labeled with core diameter/shell thickness in nanometers.

041121_1_037604jbo4.jpg

Figures 5a and 5b are analogous to Figs. 4a and 4b, but here nanoshells are added to the microsphere-based turbid phantom and intensity gains are computed relative to the phantom alone. Average A-scan data are shown in Fig. 5a. Average intensity gains shown in Fig. 5b vary from 2 to 7dB , in a monotonic relationship with the predicted μback . The level of intensity gain produced by nanoshells in the phantom material is much lower than that seen for the water-based measurements. We can interpret the intensity gains seen with this phantom as a meaningful prediction of actual intensity gains to be expected when these nanoshells are added to tissue.

Fig. 5

(a) Average A-scans of a tissue phantom and two different sizes of nanoshells in the phantom ( 5×109 nanoshells/mL). The vertical dotted lines indicate the first 100μm over which the intensity was averaged for intensity gain estimates. (b) Measured intensity gain from each size of nanoshells versus the backscattering coefficient (μback) predicted for each size of nanoshells.

041121_1_037604jbo5.jpg

The OCT signal attenuation due to nanoshells was measured in water using the intensity of the reflection from the liquid-glass interface at the bottom of a 1-mm pathlength cuvette. This interface was used to quantify attenuation because it provided a strong signal that was more easily quantifiable and had lower error levels than calculating the attenuation by the standard approach of fitting the decay curve. Given the relatively small variations in OCT signal attenuation with nanoshell parameters, this strong signal was necessary to provide an accurate illustration of changes in attenuation. The drop in this intensity when the nanoshells were added to water provides the attenuation value, as shown in Fig. 6a . Note that in this graph, the attenuation in the turbid phantom is much greater than that of the nanoshells, as indicated both by the rate of signal decay as well as the size of the second peak (or lack thereof, in the phantom case). The lack of a first order decay in the phantom indicates a significant contribution of multiply scattered light.29 Figure 6b shows that the measured attenuation in water tracks the predicted extinction coefficient (μext) of nanoshells. In this graph, the order of the 213/19 and 291/8 nanoshells has reversed as compared to the intensity gain, because the thin shell of the 291/8 nanoshells leads to increased absorption. These results indicate that nanoshells with larger core and shell have a greater potential for inducing attenuation; however, this attenuation was minimal compared to that produced by the phantom.

Fig. 6

(a) Average A-scans of water, a turbid tissue phantom and nanoshells in water ( 5×109 nanoshells/mL). The cuvette pathlength is 1mm . (b) Measured attenuation from each size of nanoshells in water as a function of the nanoshell extinction coefficient (μext) predicted from theory.

041121_1_037604jbo6.jpg

A more biologically relevant evaluation of attenuation was performed by measuring changes due to nanoshells in a turbid phantom. However, due to the high attenuation of the phantom—and subsequent absence of a significant sample-glass interface peak—it was not possible to use the same setup as for the attenuation measurements in water. Two changes were made to our approach in order to enable a measurable peak: (1) the pathlength of the sample region was decreased to 0.5mm , and (2) a mirror was used as the bottom surface of the two-piece cuvette. This setup allowed us to observe that the attenuation of the phantom alone is 54dB (relative to water). Figure 7 shows the additional attenuation caused by each type of nanoshells in the phantom. The trend in this graph is similar to that seen in Fig. 6b, though the level of attenuation is reduced.

Fig. 7

Measured incremental attenuation due to several sizes of nanoshells ( 5×109 nanoshells/mL) in a turbid tissue phantom as a function of the nanoshell extinction coefficient (μext) predicted from theory. The cuvette pathlength used here is 0.5mm .

041121_1_037604jbo7.jpg

Optimization of nanoshell-enhanced OCT will require a quantitative understanding of the influence of nanoshell concentration. To address this issue, we measured the intensity gain from three types of nanoshells at various concentrations in a scattering phantom. The nanoshell geometries chosen for this analysis—126/15, 213/19, 291/25—provide insight into the range of OCT signal intensity gains that can be produced for a given concentration. The intensity gain was determined in a similar manner as for Figs. 4b and 5b, by computing the average intensity over the first 100μm of an average A-scan. Results in Fig. 8 show a strong increase in intensity gain with concentration for each of the three nanoshells. For all nanoshell types, the intensity gain did not rise appreciably above the noise floor for concentrations less than 5×108 nanoshells/mL. As expected, the 291/25 nanoshells produced the greatest intensity gain at all concentrations over which signals were above the noise floor. Setting the threshold for detection at a level slightly above the noise floor (2dB) , the minimum concentrations required to produce threshold-level intensity gains were 1010 , 2×109 , and 109 nanoshells/mL for the 126/15, 213/19, and 291/25 geometries, respectively. At any particular concentration level, the trends in intensity gain as a function of nanoshell type are in agreement with results in Figs. 4b and 5b.

Fig. 8

Measured intensity gain as a function of nanoshell type and concentration. Data are shown for three different nanoshell types added to a turbid tissue phantom. Maximum standard deviation (n=3) was 0.8dB . Error bars are not shown for clarity.

041121_1_037604jbo8.jpg

4.

Discussion

Prior studies have established the ability of nanoshells to provide improved contrast during OCT imaging of biological tissues. This work, however, represents the first parametric experimental investigation of the factors that influence signal enhancement by nanoshells. The in vitro phantom measurements performed in this study provide unique insights into the OCT imaging characteristics of nanoshells. With this information, it is now possible to begin a rigorous analysis of the issues related to optimization of nanoshell parameters for in vivo diagnosis.

Our experimental results provide evidence that increases in core diameter over the range of 126 to 291nm and shell sizes over the range of 8 to 25nm can produce a significant increase in OCT signal intensity at 1310nm . The turbid phantom alone generated an intensity level of about 20dB above the noise floor. As indicated by Fig. 5b, 5-dB intensity gains at a nanoshell concentration of 5×109 nanoshells/mL were possible by increasing core diameter (126/26 to 291/25) or shell thickness (291/8 to 291/25). The highest intensity gain generated in a phantom was a value of 9dB (Fig. 8). This gain was produced for the 291/25 nanoshells at a concentration of 1010 nanoshells/mL.

While these results indicate that achieving large increases in the OCT signal necessitates a large core diameter and shell thickness, these same attributes can also result in greater signal attenuation. Our experimental results indicate that for all but the nanoshells with the largest overall sizes, attenuation was relatively minimal. The 291/25 nanoshells at a concentration of 5×109 nanoshells/mL produced nearly 6dB of attenuation over a 0.5-mm pathlength. Although 6dB is relatively small compared to the 54dB of attenuation produced by the turbid phantom, nanoshell attenuation might become significant over greater depths, particularly in tissues with low attenuation levels and for cases in which nanoshell concentrations are significantly greater than those investigated here. It is worthwhile noting that, according to the calculated optical properties, the relative contribution of scattering to the total attenuation increases dramatically with core diameter and shell thickness. For example, in the case of 126/15 nanoshells, absorption is 1.5 times stronger than scattering; whereas, for the 291/25 geometry, scattering is more than 5 times as strong as absorption.

Variations in the intensity gain and attenuation as a function of shell thickness were less consistent than those seen for core diameter. For nanoshells with a core diameter of 291nm , the change in predicted μback was sixfold as shell diameter increased from 8 to 25nm and the increase in experimentally measured intensity gain in the turbid phantom was 4.4dB . However, in three cases, minimal differences were seen between results for nanoshells with the same core diameter but different shell thickness. One instance involves the intensity gain data for 126/15 and 126/26 nanoshells [Figs. 4b and 5b]. The other two cases involve attenuation data for the 126/15 and 126/26 nanoshells as well as the 291/15 and 291/25 nanoshells [Figs. 6b and 7]. The changes in backscattering and attenuation over these ranges were too small to be detected by our OCT system. Although variations in nanoshell parameters do not induce consistent, linear variations in backscattering or attenuation, the trends seen in the measurements match those from the calculated optical properties.

Prior calculations for a wavelength of 830nm indicated that scattering increases monotonically with shell thickness, but varies in an irregular, nonmonotonic manner with nanoshell core diameter.15 At 1310nm , however, theoretical results indicate monotonic, albeit nonlinear, increases in scattering efficiency with both core and shell size. Not only do our results corroborate the trends expected from varying each of these two geometric parameters but also for the combined effect of the parameters as well. As indicated in Figs. 4 and 5, intensity gain increased as predicted when one parameter was held constant and the other was increased. While the predominant effect was an increase in intensity gain with core size, results clearly indicate one case in which the smaller core nanoshells (213/19) produced more signal than ones with larger core (291/8). This was due to the thicker shell of the 213/19 nanoshells and is in agreement with theory. While a fully quantitative validation of calculated optical properties will be performed in a future study, our current results provide strong corroboration with theoretically predicted trends.

One of the most important practical concerns for nanoshell-based enhancement of OCT images is the density or concentration of nanoshells that are necessary to produce a reliably detectable level of contrast. A series of measurements at different concentrations (Fig. 8) provides insight into the issue of detection threshold. As expected, nanoshells with higher backscattering efficiencies required lower concentrations to reach the detection threshold. For the most highly backscattering nanoshells (291/25), a concentration of 109 nanoshells/mL was required to produce an intensity gain of 2dB or more in the phantom.

We can compare this threshold concentration of 109 nanoshells/mL to what concentrations may be realized in cancerous tumors in vivo. First, we consider the case of nanoshells labeled to target a tumor. A tumor marker such as the epidermal growth factor receptor can be found at levels of 106 per cell on cancerous cells (versus 103 to 105 per cell on normal cells).30, 31 For the case of cells with 10μm diameter and nanoshells with 300-nm diameter, then up to 4000 nanoshells could be accommodated on the cell surface, and at least 100 receptors would be available to bind to each antibody-conjugated nanoshell. If only 10% of each cell’s surface were covered with nanoshells (400 nanoshells/cell), the nanoshell concentration would be 6.4×1011 nanoshells/mL, two orders of magnitude above the threshold concentration.

In addition to a molecularly specific mechanism of delivering nanoshells to a tumor, there is also an opportunity for intravenously injected, unlabeled nanoshells to accumulate in a tumor via passive extravasation from its leaky vasculature.32, 33, 34 However, there is limited information in the literature to precisely predict the potential concentration of unlabeled nanoshells accumulating in a tumor in this manner. Unezaki 34 found that 15% of an intravenously injected 1-mg dose of PEG-coated liposomes (198-nm mean diameter) would accumulate in 1-g tumors inoculated subcutaneously in mice. Assuming the tissue and liposome mass densities are similar to that of water (1gcm3) , then a final concentration of 3.7×1010 particles/mL accumulated in each tumor, again well above the nanoshell threshold concentration. Interestingly, there was a distinct size dependence of the extravasation, with 63-nm and 388-nm diameter liposomes showing significantly less accumulation than the 198-nm liposomes.

Given the results presented here on the variations in nanoshell optical performance, it is useful to consider what geometry is optimal for OCT imaging, or perhaps more importantly, what factors should be considered in determining whether a nanoshell is “optimal.” Because a high OCT signal level would facilitate detection, one criteria is maximum intensity gain. Within the range of parameters that were investigated in the current study, the nanoshell with the largest core (291-nm diameter) and shell (25-nm thick) provided maximum OCT signal enhancement. However, this nanoshell also produced the greatest attenuation of OCT signals. Whether or not this attenuation is significant would depend on both the distribution and concentration of the nanoshells. For the nanoshell concentrations investigated here, attenuation was much less than that of biological tissue. Therefore, the contribution to overall signal loss would likely be minimal unless the nanoshells were present in very high concentrations and/or at moderately high concentrations throughout the entire depth being scanned. On the other hand, if a highly concentrated grouping of nanoshells was present at a targeted tumor, significant localized attenuation might be produced, causing strong shadowing in the region below the tumor. This shadowing, combined with a high level of backscatter, might lead to improved detection of tumors.

While 291/25 nanoshells produced the greatest signal for a given concentration, they may not be the best choice for all OCT applications. In addition to the studies of passive extravasation discussed earlier, there is further evidence that the pharmacokinetics of nanoshells is dependent on particle size—specifically, larger particles may suffer from limited or slow biodistribution and clearance.35 Furthermore, if the amount of nanoshells delivered is limited by volume rather than particle concentration (e.g., due to safety concerns), smaller diameter nanoshells may actually be more effective at producing high OCT signals. Results in Fig. 8 indicate that the concentration of nanoshells required to achieve an intensity gain of about 4dB using 291/25 nanoshells is one-fifth that required when using 126/15 nanoshells. However, the volume of a 291/15 nanoshell is about 10 times greater than that of a 126/15 nanoshell. Therefore, 126/15 nanoshells—the smallest nanoshells tested in this study—would likely produce more signal per nanoshell volume.

5.

Conclusion

The influence of nanoshell geometry and concentration on OCT signal enhancement in turbid media was investigated through in vitro OCT measurements. Nanoshells with core diameters of 126 to 291nm and shell thicknesses of 8 to 25nm were studied in water and turbid tissue phantoms. Monotonic increases of up to 8dB in OCT intensity gain and 6dB in signal attenuation were found with increasing core and shell size for concentrations of 5×109 nanoshells/mL. These gains were not linear with either geometric parameter yet were qualitatively consistent with calculated optical properties. Results indicate that a concentration of 109 nanoshells/mL may be needed to provide meaningful (2dB) signal enhancement in a relatively homogeneous turbid medium. These results help elucidate a number of basic issues relevant to optimization of nanoshell parameters for reflectance-based diagnostics and will facilitate future studies toward the development of a viable molecular OCT imaging approach.

Acknowledgments

The authors acknowledge support from the Welch Foundation, Beckman Foundation, and the National Science Foundation Center for Biological and Environmental Nanotechnology at Rice. We also acknowledge Jeff Bush of Optiphase, Inc. for many helpful discussions on configuring and characterizing our autocorrelator.

The opinions and conclusions stated in this paper are those of the authors and do not represent the official position of the U.S. Food and Drug Administration. The mention of commercial products, their sources, or their use in connection with material reported here is not to be construed as either an actual or implied endorsement of such products by the U.S. Food and Drug Administration.

References

1. 

D. Huang, E. A. Swanson, C. P. Lin, J. S. Schuman, W. G. Stinson, W. Chang, M. R. Hee, T. Flotte, K. Gregory, C. A. Puliafito, and J. G. Fujimoto, “Optical coherence tomography,” Science, 254 1178 –1181 (1991). 0036-8075 Google Scholar

2. 

J. K. Barton, J. B. Hoying, and C. J. Sullivan, “Use of microbubbles as an optical coherence tomography contrast agent,” Acad. Radiol., 9 S52 –S55 (2002). 1076-6332 Google Scholar

3. 

T. M. Lee, A. L. Oldenburg, S. Sitafalwalla, D. L. Marks, W. Luo, F. J. Toublan, K. S. Suslick, and S. A. Boppart, “Engineered microsphere contrast agents for optical coherence tomography,” Opt. Lett., 28 (17), 1546 –1548 (2003). 0146-9592 Google Scholar

4. 

C. Xu, J. Ye, D. L. Marks, and S. A. Boppart, “Near-infrared dyes as contrast-enhancing agents for spectroscopic optical coherence tomography,” Opt. Lett., 29 (14), 1647 –1649 (2004). https://doi.org/10.1364/OL.29.001647 0146-9592 Google Scholar

5. 

Y. Yang, P. O. Bagnaninchi, S. C. Whiteman, D. G. van Pittius, A. J. El Haj, M. A. Spiteri, and R. K. Wang, “A naturally occurring contrast agent for OCT imaging of smokers’ lung,” J. Phys. D, 38 2590 –2596 (2005). 0022-3727 Google Scholar

6. 

A. L. Oldenburg, J. R. Gunther, and S. A. Boppart, “Imaging magnetically labeled cells with magnetomotive optical coherence tomography,” Opt. Lett., 30 (7), 747 –749 (2005). https://doi.org/10.1364/OL.30.000747 0146-9592 Google Scholar

7. 

C. Yang, “Molecular contrast optical coherence tomography: a review,” Photochem. Photobiol., 81 215 –237 (2005). https://doi.org/10.1562/2004-08-06-IR-266.1 0031-8655 Google Scholar

8. 

S. A. Boppart, A. L. Oldenburg, C. Xu, and D. L. Marks, “Optical probes and techniques for molecular contrast enhancement in coherence imaging,” J. Biomed. Opt., 10 (4), 41208 (2005). 1083-3668 Google Scholar

9. 

K. D. Rao, M. A. Choma, S. Yazdanfar, A. M. Rollins, and J. A. Izatt, “Molecular contrast in optical coherence tomography by use of a pump-probe technique,” Opt. Lett., 28 (5), 340 –342 (2003). https://doi.org/10.1038/425340a 0146-9592 Google Scholar

10. 

C. Yang, L. E. McGuckin, J. D. Simon, M. A. Choma, B. E. Applegate, and J. A. Izatt, “Spectral triangulation molecular contrast optical coherence tomography with indocyanine green as the contrast agent,” Opt. Lett., 29 (17), 2016 –2018 (2004). https://doi.org/10.1364/OL.29.002016 0146-9592 Google Scholar

11. 

C. Yang, M. A. Choma, L. E. Lamb, J. D. Simon, and J. A. Izatt, “Protein-based molecular contrast optical coherence tomography with phytochrome as the contrast agent,” Opt. Lett., 29 (12), 1396 –1398 (2004). https://doi.org/10.1364/OL.29.001396 0146-9592 Google Scholar

12. 

J. S. Bredfeldt, C. V. Vinegoni, D. L. Marks, and S. A. Boppart, “Molecularly sensitive optical coherence tomography,” Opt. Lett., 30 (5), 495 –497 (2005). https://doi.org/10.1364/OL.30.000495 0146-9592 Google Scholar

13. 

H. Cang, T. Sun, Z. Li, J. Chen, B. J. Wiley, Y. Xia, and X. Li, “Gold nanocages as contrast agents for spectroscopic optical coherence tomography,” Opt. Lett., 30 (22), 3048 –3050 (2005). https://doi.org/10.1364/OL.30.003048 0146-9592 Google Scholar

14. 

K. Sokolov, M. Follen, J. Aaron, I. Pavlova, A. Malpica, R. Lotan, and R. Richards-Kortum, “Real-time vital optical imaging of precancer using anti-epidermal growth factor receptor antibodies conjugated to gold nanoparticles,” Cancer Res., 63 1999 –2004 (2003). 0008-5472 Google Scholar

15. 

C. H. Loo, A. W. H. Lin, M. Lee, J. K. Barton, N. J. Halas, J. L. West, and R. A. Drezek, “Nanoshell-enabled photonics-based imaging and therapy of cancer,” Technol. Cancer Res. Treat., 3 33 –40 (2004). 1533-0346 Google Scholar

16. 

K. Chen, Y. Liu, G. Ameer, and V. Backman, “Optimal design of structured nanospheres for ultrasharp light-scattering resonances as molecular imaging multilabels,” J. Biomed. Opt., 10 024005 (2005). https://doi.org/10.1117/1.1899684 1083-3668 Google Scholar

17. 

A. W. H. Lin, N. A. Lewinski, J. L. West, N. J. Halas, and R. A. Drezek, “Optically tunable nanoparticle contrast agents for early cancer detection: model-based analysis of gold nanoshells,” J. Biomed. Opt., 10 064035 (2005). https://doi.org/10.1117/1.2141825 1083-3668 Google Scholar

18. 

J. Chen, F. Saeki, B. J. Wiley, H. Cang, M. J. Cobb, Z. Y. Li, L. Au, H. Zhang, M. B. Kimmey, X. Li, and Y. Xia, “Gold nanocages: bioconjugation and their potential use as optical imaging contrast agents,” Nano Lett., 5 (3), 473 –477 (2005). https://doi.org/10.1021/nl047950t 1530-6984 Google Scholar

19. 

C. Loo, L. Hirsch, M. Lee, E. Chang, J. L. West, N. J. Halas, and R. A. Drezek, “Gold nanoshell bioconjugates for molecular imaging in living cells,” Opt. Lett., 30 (9), 1012 –1014 (2005). https://doi.org/10.1364/OL.30.001012 0146-9592 Google Scholar

20. 

D. Sarkar and N. J. Halas, “General vector basis function solution of Maxwell’s equations,” Phys. Rev. E, 56 1102 –1112 (1997). https://doi.org/10.1103/PhysRevE.56.1102 1063-651X Google Scholar

21. 

S. J. Oldenburg, “Light scattering from gold nanoshells,” Rice University, (1999). Google Scholar

22. 

A. L. Aden and M. Kerker, “Scattering of electromagnetic waves from two concentric spheres,” J. Appl. Phys., 22 1242 –1246 (1951). https://doi.org/10.1063/1.1699834 0021-8979 Google Scholar

23. 

C. F. Bohren and D. R. Huffman, Absorption and Scattering of Light by Small Particles, Wiley, New York (1983). Google Scholar

24. 

S. L. Westcott, J. B. Jackson, C. Radloff, and N. J. Halas, “Relative contributions to the plasmon lineshape of metal nanoshells,” Phys. Rev. B, 66 155431 (2002). https://doi.org/10.1103/PhysRevB.66.155431 0163-1829 Google Scholar

25. 

D. G. Duff, A. Baiker, and P. P. Edwards, “New hydrosol of gold clusters. 1. Formation and particle size variation,” Langmuir, 9 2301 –2309 (1993). https://doi.org/10.1021/la00033a010 0743-7463 Google Scholar

26. 

J. Bush, P. Davis, and M. A. Marcus, “All-fiber optic coherence domain interferometric techniques,” Proc. SPIE, 4204 71 –80 (2000). 0277-786X Google Scholar

27. 

U. Sharma, N. Fried, and J. Kang, “All-fiber common-path optical coherence tomography: sensitivity optimization and system analysis,” IEEE J. Sel. Top. Quantum Electron., 11 (4), 799 –805 (2005). 1077-260X Google Scholar

28. 

T. Troy and S. N. Thennadil, “Optical properties of human skin in the near infrared wavelength range of 1000 to 2200nm,” J. Biomed. Opt., 6 (2), 167 –176 (2001). https://doi.org/10.1117/1.1344191 1083-3668 Google Scholar

29. 

R. K. Wang, “Signal degradation by multiple scattering in optical coherence tomography of dense tissue: a Monte Carlo study towards optical clearing of biotissues,” Phys. Med. Biol., 47 2281 –2299 (2002). https://doi.org/10.1088/0031-9155/47/13/307 0031-9155 Google Scholar

30. 

T. T. Kwok and R. M. Sutherland, “Differences in EGF related radiosensitisation of human squamous carcinoma cells with high and low numbers of EGF receptors,” Br. J. Cancer, 64 (2), 251 –254 (1991). 0007-0920 Google Scholar

31. 

G. Carpenter, “Receptors for epidermal growth factor and other polypeptide mitogens,” Annu. Rev. Biochem., 56 881 –914 (1987). 0066-4154 Google Scholar

32. 

H. Maeda, J. Fang, T. Inutsuka, and Y. Kitamoto, “Vascular permeability enhancement in solid tumor: various factors, mechanisms involved and its implications,” Int. J. Immunopharmacol, 3 319 –328 (2003). 0192-0561 Google Scholar

33. 

O. Ishida, K. Maruyama, K. Sasaki, and M. Iwatsuru, “Size dependent extravasation and interstitial localization of polyethyleneglycol liposomes in solid tumor-bearing mice,” Int. J. Pharm., 190 49 –56 (1999). 0378-5173 Google Scholar

34. 

S. Unezaki, K. Maruyama, J. Hosoda, I. Nagae, Y. Koyanagi, M. Nakata, O. Ishida, M. Iwatsuru, and S. Tsuchiya, “Direct measurement of the extravasation of polyethyleneglycol-coated liposomes into solid tumor tissue by in vivo fluorescence microscopy,” Int. J. Pharm., 144 11 –17 (1996). 0378-5173 Google Scholar

35. 

G. L. McIntire, E. R. Bacon, K. J. Illig, S. B. Coffey, B. Singh, G. Bessin, M. T. Shore, and G. L. Wolf, “Time course of nodal enhancement with CT X-ray nanoparticle contrast agents: effect of particle size and chemical structure,” Invest. Radiol., 35 (2), 91 –96 (2000). 0020-9996 Google Scholar
©(2006) Society of Photo-Optical Instrumentation Engineers (SPIE)
Anant Agrawal, Stanley Huang, Alex Wei Han Lin, Min-Ho Lee, Jennifer Kehlet Barton, Rebekah Anna Drezek, and Joshua Pfefer "Quantitative evaluation of optical coherence tomography signal enhancement with gold nanoshells," Journal of Biomedical Optics 11(4), 041121 (1 July 2006). https://doi.org/10.1117/1.2339071
Published: 1 July 2006
Lens.org Logo
CITATIONS
Cited by 109 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Optical coherence tomography

Signal attenuation

Backscatter

Gold

Tissues

Scattering

Tumors

Back to Top