20 March 2013 Spectroscopic imaging of serum proteins using quantum cascade lasers
Anadi Mukherjee, Quentin Bylund, Manu Prasanna, Yotam Margalit, Tarik Tihan
Author Affiliations +
Abstract
First measurements of biomedical imaging using quantum cascade lasers (QCL) are presented. We report spectroscopic imaging of serum proteins using QCLs as an example for monitoring surface biocontamination. We found that dry smears of human serum can be spectroscopically imaged, identified, and quantified with high sensitivity and specificity. The core parts of the imaging platform consist of optically multiplexing three QCLs and an uncooled microbolometer camera. We show imaging of human serum proteins at 6.1, 9.25, and 9.5 μm QCLs with high sensitivity and specificity. The sensitivity limit of 3  μg/cm2 of the human serum spot was measured at an S/N=3 .The specificity of human serum detection was measured at 99% probability at a threshold of 77  μg/cm 2 . We anticipate our imaging technique to be a starting point for more sophisticated biomolecular diagnostic applications.

1.

Introduction

Surface biocontamination is a major public health hazard that results in contaminated foods, drugs, cross-contamination of diseases in clinics/hospitals and public places such as restaurants.1 An image-based device for full-surface screening of biocontamination can assure higher-quality healthcare for our citizens. Trace detection techniques such as mass spectrometry,2,3 Fourier transform infrared spectroscopy (FTIR),4 and chromatography5 require swabbing a small sample area for testing. Although sensitive, these procedures are time-consuming, expensive, and do not cover 100% of the area that needs to be monitored regularly. To reduce public health hazards due to surface biocontamination significantly, a portable imaging system is needed that can detect, identify, quantify, and map out surface biocontamination on any amount of surface area with high sensitivity and specificity. Laser-based chemical imaging techniques explored so far are Raman6 and near-infrared (NIR)7,8 scattering. Fundamental absorption cross-section (1017cm2mol1)>overtone absorption cross-section (<1019cm2mol1)>Raman scattering cross-section (1029cm2mol1).9,10 Here we show that an efficient tag-less biomolecular imaging device can be made out of midwavelength infrared (MIR) lasers called quantum cascade lasers (QCL) and long wavelength infrared (LWIR) uncooled microbolometer cameras. We found that dry smears of human serum can be spectroscopically imaged, identified, and quantified with high sensitivity and specificity. We anticipate our imaging technique to be a starting point for more sophisticated biomolecular diagnostic applications.

2.

Methods, Results and Discussion

When a MIR laser beam is tuned to an absorption peak of biomolecules of interest (e.g., Amide I peak of proteins), the beam is absorbed, and the scattered image will appear dark. A three-dimensional (3-D) stack of MIR images at different wavelengths (spectral hypercube) contains the entire biomolecular signature needed for diagnosis. Molecular recognition is most sensitive and selective at the absorption peaks. Images at the strongest absorption peak will be used for sensitivity measurement, and images at other wavelengths will be used to quantify specificity of detection via receiver operation characteristic (ROC) curves. Here we show the results of identifying human serum at a distance of 50 cm using our compact MIR imaging spectroscopy (MIRIS) setup.

Figure 1(a) shows the MIR imaging device with optically multiplexed 3 QCLs and a microbolometer camera. The home-built software operates the lasers, acquire and process images. Figure 1(b) shows the device details with three multiplexed QCLs, camera, digital to analog converter (DAC) board, and other optoelectronics. Image-based detection of explosives in the MIR have been reported via optothermal10,11 and MIRIS12,13 techniques. Optically multiplexed QCLs as reported here, are preferred to one broadly tunable1013 QCL because of wider frequency bandwidth availability (>3000cm1) with higher optical powers. The three QCLs (6.1, 9.25, and 9.5 μm) were multiplexed14 using a Galvo-mirror from Cambridge Technologies. The multiplexed beam was expanded to 1.5cm diameter on an aluminum plate 50 cm away from the device. The QCLs at were DFB devices while the QCL at 6.1 μm was a FP (Fabry-Perot) device. The QCLs obtained from www.atoptics.com were thermally cooled at 17°C. A DAC board and Linux OS were used for full device control.

Fig. 1

Quantum cascade lasers (QCL) based MIR imaging system: (a) the device for QCL-based spectroscopic biomedical imaging. The software operate lasers, acquire and process images, (b) detail of the imaging device showing optically multiplexed three QCLs at wavelengths 9.25, 9.5, and 6.1 μm camera etc.

JBO_18_3_036011_f001.png

Figure 2(a) shows the dried spot of random human serum on a CaF2 slide. The serum spot is invisible when illuminated by visible light. The outline of the QCL beams is shown. Dried serum smears were produced by using a hole punched onto nonabsorbing tape as a mask for pattern transfer. 5 μl of serum was smeared onto the hole of the tape and dried. The tape was then removed, leaving behind the dried serum spot on the slide. This way, dried serum smears with repeatable dimensions were produced. Diffuse scattering at each QCL frequency from the aluminum plate placed behind the slide were recorded by the microbolometer camera obtained from www.infraredcamerasinc.com.

Fig. 2

Spectroscopic image detection of human serum on CaF2 slide: (a) shows almost invisible serum spot on CaF2 slide and the outline of illuminating QCL laser beam; (b) absorbance of the serum spot at 6.1 μm in 3-D corresponding to dark spot in shown in (c), FTIR spectrum recorded on the human serum and background subtracted MIR images of the serum spot taken by the camera under three QCL illumination at 9.5 μm (1052cm1), 9.25 μm (1081cm1), and 6.1 μm (1639cm1).

JBO_18_3_036011_f002.png

Figure 2(b) shows the absorbance of the serum spot in 3-D when exposed to 6.1 μm QCL beam. The incident beam intensity normalized serum protein signal show strong sensitivity achievable in MIRIS based detection. This 3-D representation of absorbance of serum smear at 6.1 μm is shown in transmission (scatter from aluminum plate behind sample) in Fig. 2(c). The aluminum plate (5×10 in) used in the measurements was purchased off the shelf from a local hardware store and had no polish, only matte finish. Figure 2(c) also shows FTIR spectrum of the human serum recorded by Shimadzu IRAffinity-1 ATR-FTIR spectrometer. The FTIR data was stored after averaging over 40 spectra, with 4cm1 resolution and Happ-Genzel apodization. Three QCLs (6.1, 9.25, and 9.5 μm) are marked by arrows on the FTIR spectrum. The QCL at 6.1 μm (1639cm1) being close to the Amide-I peak of proteins was chosen for high sensitivity protein imaging. The other two QCLs at 9.25 μm (1081cm1) and 9.5 μm (1053cm1) were around the glucose peak of the serum. Colinear QCL beams of CW power 11mW of each beam were turned on one at a time for illuminating the sample area on CaF2 slide. The background-subtracted images of the dried serum spot at each laser wavelength are shown. The dried serum spot on CaF2 slide that was invisible to visible light was clearly visible when illuminated by the 6.1 μm QCL beam. The serum spot looks dark due to the strong absorption of 6.1 μm QCL beam by Amide-I proteins while partial transparency (reduced absorption) is seen at the Glucose wavelengths (QCLs at 9.25 and 9.5 μm) as expected from the FTIR spectrum of serum. We see a good correspondence between the absorption values of FTIR spectrum with that of the pixel values of the recorded MIR images at the three QCL frequencies. The absorption ratio of Amide-I/Glucose 3.12 in FTIR spectrum is close to the value 3.2 measured in the central pixel values of the serum spot images. The dark annular ring of the serum spot at the Glucose wavelengths is due to thicker film at the edges than at the central region. We have studied 10 different human serum samples using MIRIS technique and all of them show the same behavior as the random one shown in this report. We found a standard error of 5% around the signal-to-noise ratio of the random sample result shown in Fig. 2(c). The fringe pattern seen on the images are due to interference of QCL beams from the front and back surfaces of the slide. These fringe patterns are expected to be different on different surface materials on which the smear may be found (e.g., if the smear was directly on the aluminum plate, the fringes due to thin film effect of the smear and laser speckle patterns would influence the limit of detection). Surface quality of the substrate will also play a role in the diffuse scattering signal.12,13 Laser despeckling can be used to get rid of these fringes and laser speckle that increase limit of detection13 and reduce spatial resolution of imaging.

Figure 3 shows sensitivity of MIR image-based serum protein detection. Figure 3(a) shows the scatter image of 6.1 μm QCL beam (no serum) tuned to Amide-I protein peak. This beam profile was incident on dried serum smears at different dilutions to test the limit-of-detection (LOD) of our imaging system. Interference fringes seen on the incident beam is due to front and back surface reflections on the CaF2 slide, and this was the source of most of the noise. The images shown was averaged over 1 s (i.e., 50 frames of the camera and the power stability of the lasers were <2%). The CaF2 slides were polished on both sides by the manufacturer, which turned out to have a narrow wedge angle resulting in the interference fringes observed. The fringes are correlated to the 2-mm slide thickness and refractive index changes of CaF2 at the QCL wavelengths due to normal dispersion. The refractive index of at 9 μm is 1.324, and at 6 μm is 1.383, and hence we see slightly narrower fringes at the shorter wavelength images. Figure 3(b) show the sequence of scatter images of the QCL beam at 6.1 μm for different serum dilutions starting with pure serum of 403μg/cm2 surface density. The surface density measurements shown in Fig. 3 included all solid biomolecular components in human serum, not just proteins (i.e., it includes proteins, nucleic acids, phospholipids, sugars, membrane lipids, fatty acids etc.,). Average protein content in human serum is 70g/L. For 5 μL of pure serum, we can expect around 350 μg of total proteins (Amide I, II, and II) in the measured 403μg/cm2 containing all biomolecular components in human serum. Since the QCL at 6.1 μm was tuned to Amide I peak, we were measuring dominantly Amide I, some Amide II, and a little of Amide III proteins due to absorption tails at this QCL wavelength. At each step of dilution, an equal volume of phosphate buffer was added to the last serum sample and dried in identical manner. No image processing was used other than background subtraction. The decrease of contrast on serum detection versus decrease of surface density due to dilution of serum can be clearly seen. At the lowest total surface density of 3μg/cm2, the serum spot was still visible with an average S/N=3. Considering the fact that we were actually sensing only proteins out of all biomolecular components in the total surface density, our measured LOD of 3μg/cm2 was a conservative estimate of total proteins in human serum. To the best of our knowledge, this is the lowest limit of detection reported so far13 in any noninvasive image-based molecular recognition system. With despeckled laser beams and advanced image-processing techniques, we anticipate achieving an LOD of several 100 s of ng/cm2. Such high sensitivity in image-based detection of biomolecules may perhaps enable quantitative stain-free pathology in the future.

Fig. 3

Sensitivity of MIR image detection of human serum: (a) image of 6.1 μm QCL beam scattered off the aluminum plate placed behind the CaF2 slide with no serum smear; (b) background subtracted MIRIS images of serum and diluted serums on CaF2 slide taken near the Amide I peak of 6.1 μm (1639cm1). The images of pure (403μg/cm2) and diluted serums show the decrease of contrast with diluted serum spots. The LOD of 3μg/cm2 was achieved without the use of laser despeckling or advanced image processing.

JBO_18_3_036011_f003.png

Figure 4 shows the specificity of MIR image-based serum protein detection. Dried spots of serum look-alikes on the same slide (CaF2) are indistinguishable under visible light illumination. Figure 4(a) shows MIR image of lemon ice tea that looks like serum in visible light but is clearly distinguishable using proper choice of QCL lasers. Also, serum protein may be found in myriads of surfaces that may have absorptions in the protein region. Such interferent rejection can be done following gas detection15 using QCLs. Specificity is quantified as probability of detection (PD) or false alarm rates probability of false alarm (PFA) from ROC curve. The ROC curve calculations from our serum protein measurements on blank CaF2 slide is shown in Fig. 4 following procedure reported in QCL based gas detection.15

Fig. 4

Specificity of image detection of human serum: (a) lemon ice tea as a serum look-alike target; (b) ROC curve demonstrating PFA of 102 (PD of 99%) of human serum detection at a threshold of 77μg/cm2.

JBO_18_3_036011_f004.png

3.

Conclusion

As a simple demonstration of the first biomedical imaging using quantum-cascade lasers, we spectroscopically imaged trace amounts of human serum with high sensitivity and specificity. Using an optically multiplexed platform of three quantum cascade lasers and a microbolometer camera, we detected trace amounts of proteins in human serum as low as 3μg/cm2. High specificity of protein detection in human serum was demonstrated by the spectroscopic images of lemon ice tea taken as an example of a serum look-alike substance. Specificity was quantified via ROC curve showing PFA of 102 (99% PD) for human serum detection at a threshold of 77μg/cm2. By a proper choice of QCL wavelengths and/or multiplexing more lasers, we would be able to distinguish between different biomaterial sources containing proteins. This image-based biocontamination sensor can be applied in the areas of security from biohazards, food safety, cross contamination in clinical ambiance, etc. It is also a preliminary example showing the potential of MIRIS-based technology toward most advanced label-free, noninvasive medical diagnostic applications that are well known in the biomedical industry.

Acknowledgments

We gratefully acknowledge laboratory support and associated help from Department of Pathology, UCSF; help we received from Mariano Troccoli of Adtech Optics regarding QCLs, Gary Strahan of Infrared Cameras Inc., regarding microbolometer camera and many helpful discussions with Biswaroop Mukherjee throughout the course of this project.

References

1. 

“Cleanrooms and associated controlled environments—biocontamination control—Part 2: Evaluation and Interpretation of Biocontamination data,” (2010) http://www.ihs.com Google Scholar

2. 

P. P. HurtadoP. B. O’Connor, “Differentiation of isometric amino acid residues in proteins and peptides using mass spectrometry,” Mass Spectrom. Rev., 31 (6), 609 –625 (2012). http://dx.doi.org/10.1002/mas.v31.6 MSRVD3 0277-7037 Google Scholar

3. 

M. Bantscheffet al., “Quantitative mass spectrometry in proteomics: a critical review,” Anal. Bioanal. Chem., 389 (4), 1017 –1031 (2007). http://dx.doi.org/10.1007/s00216-007-1486-6 ABCNBP 1618-2642 Google Scholar

4. 

A. Barth, “The infrared absorption of amino acid side chains,” Rev. Prog. Biophys. Mol. Biol., 74 (3–5), 141 –173 (2000). http://dx.doi.org/10.1016/S0079-6107(00)00021-3 PBIMAC 0079-6107 Google Scholar

5. 

B. Shushan, “A review of clinical diagnostic applications of liquid chromatography-tandem mass spectrometry,” Mass Spectrom. Rev., 29 (6), 930 –944 (2010). http://dx.doi.org/10.1002/mas.v29:6 MSRVD3 0277-7037 Google Scholar

6. 

B. Shushan, “Label-free imaging of biomolecules in food products using Raman microscopy,” J. Biomed. Opt., 16 (2), 021118 (2011). http://dx.doi.org/10.1117/1.3516591 JBOPFO 1083-3668 Google Scholar

7. 

U. Mahmood, “Near infrared optical applications in molecular imaging,” IEEE Eng. Med. Bio., 23 (4), 58 –66 (2004). http://dx.doi.org/10.1109/MEMB.2004.1337950 IEMBDE 0739-5175 Google Scholar

8. 

G. Reich, “Near-infrared spectroscopy and imaging: basic principles and pharmaceutical applications,” Adv. Drug Deliv. Rev., 57 (8), 1109 –1143 (2005). http://dx.doi.org/10.1016/j.addr.2005.01.020 ADDREP 0169-409X Google Scholar

9. 

W. Demtroder, “Laser spectroscopy: basic concepts and instrumentation,” Springer Series in Chemical Physics, Springer-Verlag, Berlin Heidelberg (1982). Google Scholar

10. 

A. Mukherjeeet al., “Standoff detection of explosive substances at distances up to 150 meters,” Appl. Opt., 49 (11), 2072 –2078 (2010). http://dx.doi.org/10.1364/AO.49.002072 APOPAI 0003-6935 Google Scholar

11. 

C. A. Kendzioraet al., “Infrared photothermal imaging for standoff detection applications,” Proc. SPIE, 8373 83732H (2012). http://dx.doi.org/10.1117/12.918594 PSISDG 0277-786X Google Scholar

12. 

B. E. BernackiM. C. Phillips, “Standoff hyperspectral imaging of explosives residues using broadly tunable external cavity quantum cascade laser illumination,” Proc. SPIE, 7665 766501 (2010). http://dx.doi.org/10.1117/12.849543 PSISDG 0277-786X Google Scholar

13. 

F. Fuchset al., “Imaging standoff detection of explosives using widely tunable midinfrared quantum cascade lasers,” Opt. Eng., 49 (11), 111127 (2010). http://dx.doi.org/10.1117/1.3506195 OPEGAR 0091-3286 Google Scholar

14. 

A. Mukherjeeet al., “Optically multiplexed multi-gas detection using quantum cascade laser photoacoustic spectroscopy,” Appl. Opt., 47 (27), 4884 –4887 (2008). http://dx.doi.org/10.1364/AO.47.004884 APOPAI 0003-6935 Google Scholar

15. 

A. Mukherjeeet al., “Sub-parts-per-billion level detection of dimethyl methyl phosphonate (DMMP) by quantum cascade laser photoacoustic spectroscopy,” Appl. Opt., 47 (10), 1543 –1448 (2008). http://dx.doi.org/10.1364/AO.47.001543 APOPAI 0003-6935 Google Scholar
© 2013 Society of Photo-Optical Instrumentation Engineers (SPIE) 0091-3286/2013/$25.00 © 2013 SPIE
Anadi Mukherjee, Quentin Bylund, Manu Prasanna, Yotam Margalit, and Tarik Tihan "Spectroscopic imaging of serum proteins using quantum cascade lasers," Journal of Biomedical Optics 18(3), 036011 (20 March 2013). https://doi.org/10.1117/1.JBO.18.3.036011
Published: 20 March 2013
Lens.org Logo
CITATIONS
Cited by 3 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Quantum cascade lasers

Proteins

Imaging spectroscopy

Spectroscopy

Absorption

Cameras

Multiplexing

Back to Top