1.IntroductionNon-invasive functional brain mapping is a technique that allows the localization of functional areas of the patient’s brain. This technique is used during brain tumor resection surgery to indicate to the neurosurgeon the cortical tissues that should not be removed to avoid motor, speech, and cognitive impairments. Functional magnetic resonance imaging (fMRI)1 is the gold standard for preoperative identification of the patient’s functional brain areas. However, after the patient’s craniotomy, neuronavigation is rendered unreliable when brain shift invalidates the patient-to-image registration and, therefore, does not allow a precise identification of the functional areas.2 To avoid localization errors, intraoperative MRI has been suggested,3 but it is costly and time-consuming, and above all, intraoperative MRI is available only in a very restricted number of neurosurgical centers.4 During neurosurgery, electrical brain stimulation (EBS) is the gold standard for intraoperative functional brain mapping,5 but this technique is mainly limited by its low spatial resolution (6) and has the potential risk to trigger epileptic seizures. This technique allows robust and reliable detection of brain functions but could be traumatic for the patient. Indeed, when testing cognitive functions such as speech, patients are awake, and EBS provokes transient disturbances in the patient that inhibit the function.5 Optical imaging is an excellent complement to EBS as this technique is contactless, non-invasive, and non-ionizing and has a low traumatic impact on the patient. Indeed using optical imaging, the task-based paradigms used to localize the brain functions are similar to those used for fMRI.7 The analysis of the light absorbance allows to monitor the brain activity (e.g., motor or sensory tasks) with quantification of chromophores in the brain cortex: the concentration changes in oxy-() and deoxygenated hemoglobin ()7–11 and cytochrome-c-oxidase (),12–14 a mitochondrial marker of metabolism. CCO is an enzyme in the mitochondria of neuronal cells and is the terminal electron acceptor in the electron transport chain.15 Total CCO concentration changes are slow and not correlated with brain activation. However, during cerebral activity, there are symmetrical variations in its oxidized and reduced states.14 Near-infrared spectroscopy (NIRS) studies showed that it is possible to resolve changes in the oxidation state of CCO using broadband spectroscopy procedures.14,16,17 CCO is not the only enzyme that takes part in the electron transport chain; other chromophores are involved such as cytochrome and (Cytb, Cytc). In in vivo NIRS studies, oxCCO changes are measured with near-infrared light. To observe and measure other chromophores, visible light is required. However, this spectral range is difficult to use in vivo due to the poor penetration of the light in the tissue. During neurosurgery, the cerebral cortex is exposed, making it possible to measure cerebral hemodynamics and metabolism at visible and near-infrared wavelengths. Robust quantification of hemodynamic and metabolism markers is complicated to perform during neurosurgery due to the critical context of the operating room, which makes the calibration of optical devices more complex. To overcome this issue, tissue-simulating objects are required for the development of medical imaging devices. These so-called “phantoms” may be used to evaluate, optimize, compare, or control imaging systems.18 Some phantom recipes based on Intralipid and blood19 and cytochrome-contained yeast20,21 have emerged to reach that goal. The purpose of these devices is to evaluate the capacity of the acquisition system to follow the variations of oxygenation of the liquid. However, it is not possible to model hemodynamic responses similar to those that occur in the brain. As these phantoms are based on Intralipid and blood, it is not possible to create heterogeneous phantoms that mimic the vascular network of the brain. For this reason, digital brain phantoms may be especially adapted to model hemodynamic and metabolic changes similar to those occurring during cerebral activity22,23 as well as 3D heterogeneous grey matter volumes.24 With wide-field imaging devices implying the use of a camera and homogeneous illumination of the cerebral cortex, the biomarkers of brain hemodynamics and metabolism are quantified with the modified Beer–Lambert law.7,13 This spectroscopic technique is however subject to quantification errors such as crosstalks14 and partial volume effect.25 The crosstalk is defined as a concentration change of a chromophore that induces a change in another chromophore.17 The partial volume effect is wavelength-dependent and refers to the impact of a focal change in the optical properties of a portion of the tissue on the surrounding tissue. To reduce the impact of the partial volume effect, the optical mean path length used in the modified Beer–Lambert law needs to be precisely estimated to consider the inhomogeneities of the optical properties.9,24 In NIRS applications, the concentration of the chromophores is usually resolved with a fitting procedure, which requires a great number of wavelengths.26 In in vivo applications, broadband spectra in the near-infrared range between 780 and 900 nm are adapted for monitoring changes in , Hb, and oxCCO due to the differences between the molar extinction spectra of , Hb, and the strong peak in the molar extinction spectra of oxCCO, as well as the deep penetration of the light into the tissue and low scattering effects. For acquiring broadband spectra for wide-field applications, hyperspectral imaging techniques need to be used.27,28 These broadband systems require complex instrumentation that implies a compromise in the spatial, spectral, and temporal resolution of the acquisitions. Thus, the number of spectral bands in these systems is often limited by the experimental devices. Hyperspectral devices reconstruct spatially and spectrally resolved images. It generates a large amount of data that could be difficult to process in real time. For these reasons, specific wavelengths have been proposed in the literature to monitor hemodynamic and metabolic changes. Bouchard et al.29 used a monochrome camera combined with two-wavelength illumination (470 and 530 nm) to monitor hemodynamic changes in a mouse model during cerebral activity. In the same way, White et al.30 used two more wavelengths (470, 530, 590, and 625 nm). Arifler et al.31 proposed an optimization procedure to identify a reduced number of wavelengths to monitor , Hb, and oxCCO within the spectral range 780 to 900 nm. Authors showed that a combination of eight wavelengths (i.e., 784, 800, 818, 835, 851, 868, 881, and 894 nm) allows a quantification error of compared with a reference measurement (broadband spectra between 780 and 900 nm). Leadley et al.32 investigated the capability of NIRS systems to monitor , Hb, and oxCCO within the spectral range of 700 to 900 nm. The authors used an analytical model based on the diffusion equation to explore the effect of wavelength choice, spectral bandwidth, and uncertainties in extinction coefficient and path length on the quantification of concentration changes. The authors showed that a wrong evaluation of the extinction coefficients leads to a poor estimation of , Hb, and oxCCO. In this specific case, NIRS systems that use fewer than 50 measurement wavelengths may not be capable of accurately measuring , Hb, and oxCCO signals. In this study, we developed a digital instrument simulator to optimize the development of hyperspectral systems for intraoperative brain mapping studies. This simulator can provide realistic modeling of the cerebral cortex and the identification of the optimal wavelengths to monitor cerebral hemodynamics (oxygenated and deoxygenated hemoglobin Hb) and metabolism (oxidized state of cytochrome and and cytochrome-c-oxidase oxCytb, oxCytc, and oxCCO). 2.Material and Methods2.1.Digital Instrument SimulatorThe digital instrument simulator is based on a realistic digital phantom of an exposed cortex computed with Monte Carlo simulations. The code is publicly available on GitHub. This simulator aims to model the light propagation in the exposed brain. It can be used to take into account the inhomogeneities of the optical properties of the exposed brain cortex for functional brain mapping studies and to reduce the quantification errors of biomarkers of brain functionality. The Monte Carlo framework is composed of several steps, see Fig. 1. First, a color image of the exposed cortex was taken during the neurosurgery, see Sec. 2.1.1. This image was segmented into three classes (i.e., grey matter and large and small blood vessels, see Sec. 2.1.2). A brain volume was then modeled, and a white Monte Carlo approach was used to estimate the partial path length, the position, and the angle of exiting packets of photons, see Sec. 2.1.3. Finally, the information of the exiting packets of photons was converted into images of diffuse reflectance and mean path length, see Sec. 2.1.4. 2.1.1.Data collectionA color image of a human brain’s exposed cortex was collected during a brain tumor resection operation at the neurological center of the Pierre Wertheimer Hospital in Bron (France). An 8-bit image was acquired with a surgical microscope Leica M530 OHX, Leica Microsystems SAS, Wetzlar, Germany ( with a resolution of ). The study was conducted according to the guidelines of the Declaration of Helsinki and approved by the local ethics committee of Lyon University Hospital (France, Gliospect: 69 HCL14-0270). The participating patients signed written consent. 2.1.2.Image segmentation and brain volume modelingThe color image was segmented into three classes (i.e., grey matter and large and small blood vessels) using morphological operations, see Fig. 2.
Once the image was segmented into three classes, we modeled the brain volume. First, the binary segmentation masks were encased in a larger image having the label of grey matter. This operation is performed to prevent a blood vessel label from appearing at the edges of the image. The objective is to avoid photon loss due to the boundary effects during the Monte Carlo simulations, see Sec. 2.1.3. The binary segmentation masks were expanded along the vertical direction ( axis) on 2 cm to convert the binary images into binary volumes, see Figs. 3(a) and 3(b). Then, we modeled the 3D blood vasculature with morphological erosion, see Fig. 3(c). The thickness of the blood vessels was calculated on the basis of the vessel’s diameter. The structuring element used for the erosion was a disk of 0-pixel diameter for (in pixels) and was increased to 1 pixel while increasing axis. The binary volumes of the three classes were finally merged together with a final isotropic resolution of , see Fig. 1. With this model of the cerebral vasculature, the blood vessels were only modeled on the surface of the volume. Deep blood vessels (which are not visible in color imaging) were not included. 2.1.3.White Monte Carlo simulationsOnce the brain volume was defined, we computed Monte Carlo simulations of light propagation in the domain using MCX34 software, see Fig. 4. A planar light source was placed on the top of the volume to homogeneously illuminate the brain volume. We set the Fresnel reflection as boundary conditions for the top and bottom faces of the volume (i.e., an exiting packet of photons is lost). Cyclic boundary conditions were applied for the sides of the volume to avoid photon loss due to the boundary effects. Each voxel of the volume included optical properties, see Table 1. Table 1Optical properties used in the white Monte Carlo simulations. λ denotes the wavelength in nanometers.
For these simulations, we implemented a white Monte Carlo approach, and the absorption coefficient was set to . With this technique, the packets of photons were not affected by the absorption. The advantage of the white Monte Carlo approach is that the absorption can be considered “a posteriori,” using the microscopic Beer–Lambert law.39 Thus, when changing the absorption parameters of the volume (i.e., to simulate a brain activation for example), the simulation does not need to be run again, speeding up the processing. In this case, only one simulation per wavelength is required, considering that the scattering properties of the medium do not change. A scattering change can be considered, but a new simulation with a new set of would need to be rerun. For each wavelength (from 400 to 1000 nm by steps of 10 nm), three outputs were stored for every packet of photons exiting the top face of the volume: the position of the exiting packet of photons , its exiting angle, and the partial path length (the length that each photon has spent in each class of the domain). 2.1.4.Image reconstructionOnce the white Monte Carlo simulations were performed (see Sec. 2.1.3), a large amount of data were generated for each wavelength (position, angle, and partial path lengths of exiting photons), see Fig. 4. For each wavelength, around 7 Gb of data needs to be processed to reconstruct the reflectance and mean path length images. With the aim of speeding up image reconstruction, a multithreaded program C++ code has been developed. To reconstruct the diffuse reflectance and mean path length images, the absorption coefficients were computed with the chemical composition of each tissue.35 These values were taken from the literature and corresponded to a nominal physiological condition,7,40–42 see Table 2. A cerebral activation will be considered later in the paper, see Sec. 2.2. With our equipment, it is not possible to determine whether the blood vessels identified by the segmentation method are arteries or veins. For this reason, we considered two cases: (1) the large and small blood vessels are arteries, and (2) the large and small blood vessels are veins. Table 2Chemical composition of the modeled tissue.
The diffuse reflectance and mean path length images can be reconstructed on a camera sensor that could be located on the tissue surface or outside the volume using a lens system. On the surface of the volume, the number of pixels is at most equal to that of the modeled surface. A spatial binning can be performed to increase the signal-to-noise ratio at the expense of spatial resolution. To reconstruct the image on the modeled camera sensor using a lens system, we used the transfer-matrix method.44 With this approach, we can consider key parameters of the optical system (focal length, working distance, size of the optics, and size of the camera sensor). A lens with a focal length of (in mm) was placed at a distance (in mm) from the surface of the tissue (along the axis ), and the sensor was located at a distance (in mm) from the lens (along the axis ). The optical system is modeled with a system matrix where and denote the translation matrices for the ray in air before and after the lens. is the lens matrix. To model the acquisition of the photon by the pixel of the camera sensor, we applied the system matrix on packets of photons exiting the tissueThe diffuse reflectance is calculated for each pixel, either on the tissue surface or on the camera sensor45,46 where is the number of packets of photons detected at the pixel , is the area (in ) of the pixel, and is the weight of the detected packet of photons defined byis the number of regions modeled in the tissue (grey matter and large and small blood vessels), (in ) is the absorption coefficient for the wavelength of the region , and is the partial path length (in mm) of the detected packet of photon that traveled in region for wavelength . The mean path length of traveled photons in the tissue is calculated in Eq. (5)46 Finally, we applied an adaptive non-local means filter on reconstructed diffuse reflectance and mean path length images for each wavelength. This operation aims to improve the signal-to-noise ratio of the Monte Carlo simulations.47 Finally, we performed a linear interpolation for each pixel of the diffuse reflectance and mean path length images on the wavelength range 400 to 1000 nm by steps of 1 nm. 2.2.Identification of the Optimal Wavelength for Hemodynamic and Metabolic MonitoringUsing the digital instrument simulator (see Sec. 2.1), we developed an optimization procedure based on a genetic algorithm (differential evolution) to identify the best wavelength combination in the visible and near-infrared range to minimize the quantification error (root mean square error in ) in , Hb, oxCCO, oxCytb, and oxCytc compared with ground truth values, see Fig. 6. 2.2.1.Model of cerebral activationData used for the optimization procedure were generated with the digital instrument simulator. A simple model was considered, which includes non-activated and activated grey matter, see Fig. 5(a). In Fig. 5(b), we modeled a cerebral activity in the activated grey matter by changing the concentration of chromophores relative to their nominal values. The modeled concentration changes that are the ground truth in the optimization procedure are indicated in the rest of the paper. A simple temporal perturbation of was performed: rest (nominal values, see Table 2) and activity (nominal value ). The values were and for and Hb,9,48 respectively, and for oxCCO,49 oxCytb, and oxCytc, mirrored by concentration changes of for redCCO, redCytb, and redCytc (the concentration of cytochromes does not change during the cerebral activity). 2.2.2.Optimization procedureThe most straightforward method to determine optimal wavelengths for cerebral monitoring is to test all possible -element combinations out of 601 wavelengths (from 400 to 1000 nm) and to identify the combination that produces the smallest estimation errors when compared with the ground truth . As the investigated spectral range is large, such an exhaustive search results in a large time complexity. Thus, we proposed to use a genetic algorithm50 to determine the optimal wavelengths through an optimization procedure, see Fig. 6. We used the differential evolution method that optimizes a problem by iteratively trying to improve a candidate solution with regard to the ground truth. We used the function differential evolution from the Python library Scipy (v1.10.0).51 The optimization procedure consists of finding the best -element combinations out of 601 wavelengths that minimize a cost function. This function takes as input a wavelength array, with values that are limited to the simulated spectral range (between 400 and 1000 nm). The cost function returns the root mean square error RMSE (in ) computed between noisy concentration changes measured with the modified Beer–Lambert law and the ground truth where is the number of iterations for the noise addition. Noise was added to the simulated intensities to identify the -wavelengths the most robust to the noise. The intensities were converted into concentration changes measured with the modified Beer–Lambert lawIn Eq. (7), is the molar extinction coefficient of the chromophore (in ), see Fig. S1 in the Supplementary Material. is the noisy attenuation change measured for , such as where is a zero-mean Gaussian noise with a standard deviation that was the same for each wavelength: . With , the diffuse reflectance simulated during rest and the signal-to-noise ratio of the instrument. This value corresponds to the experimental devices we used in a previous study,22 but this could be adapted for another experimental device. The concentration changes were obtained through matrix inversion in the least square sense.3.Results3.1.Digital Instrument SimulatorThe digital instrument simulator aimed to simulate hyperspectral images of the exposed brain cortex for 601 wavelengths (between 400 and 1000 nm). In Fig. 7, we represented the images of diffuse reflectance and mean path length at 500 and 900 nm reconstructed at the surface of the tissue considering that large and small blood vessels are arteries. The images were reconstructed with a binning, which leads to a resolution of . The red, grey, and magenta points indicate the position of points for a large blood vessel, for grey matter, and for a small blood vessel, respectively. In Fig. 8, we represented the diffuse reflectance and mean path length spectra measured for the three points identified in Fig. 7. Solid and dotted lines indicate that large and small blood vessels are considered arteries and veins, respectively. For both cases, we can observe the effect of the blood vessels on the simulated quantities. For wavelengths lower than 600 nm, the diffuse reflectance was almost on the large blood vessels; this means that hardly any photons left the large blood vessel. Diffuse reflectance values increased when measuring small blood vessels, indicating a greater contribution from outgoing photons. The higher values can be found for grey matter regions. For wavelengths higher than 600 nm, diffuse reflectance values were almost the same for grey matter and small blood vessels, and lower values can be found at the level of the large blood vessels. The mean path length spectra measured for grey matter and the small blood vessels were almost the same for all wavelengths, but lower values were obtained for the large blood vessels. We can observe that the values of the mean path length are not impacted if all vessels are considered either veins or arteries. However, we can observe differences among the diffuse reflectance spectra measured at the level of the large blood vessels for wavelengths higher than 600 nm. There is a drop in intensity in the red and near infrared for veins compared with arteries. This corresponds to the fact that veins appear darker or bluer than arteries. 3.2.Identification of the Optimal Wavelength for Hemodynamic and Metabolic MonitoringIn Fig. 9, we represented the optimal combination of 2, 4, 6, 8, and 10 wavelengths for monitoring and changes in activated grey matter. These wavelengths are represented in Fig. 9(a) by dots. We also represented the configuration of two wavelengths used by Bouchard et al.,29 the four wavelengths used by White et al.30 and the broadband spectral range between 780 and 900 nm used by Bale et al.14,17 and Arifler et al.31 The quantification errors related to these configurations are plotted in Fig. 9(b). The bars represent the RMSE averaged over 1000 noisy measurements, see Eq. (6), and the vertical line represents the standard deviation in the error. Finally, the concentration changes in and averaged over the 1000 noisy measurements are plotted in Figs. 9(c) and 9(d), respectively. Contrary to the configurations proposed by Bouchard et al. and White et al., the wavelengths identified with the optimization procedure do not include the hemoglobin isobestic point at 530 nm. The wavelengths were located in the visible and near-infrared range. Wavelengths were identified before 500 nm, where absorption predominates, between 590 and 730 nm where Hb absorption predominates and on either side of the isobestic point at 800 nm. The minimum errors in and are obtained with the combination of 10 wavelengths calculated with our optimization procedure. Among the literary configurations, White’s spectral configuration minimizes errors in and . Errors in and obtained with the optimal group of 10 wavelengths were 29% and 23% lower than those obtained with White’s spectral configuration. The mean and standard deviation of the RMSE decrease slightly as more spectral bands are considered. For all wavelengths, the standard deviation and the mean RMSE are lower than those obtained with White’s spectral configuration. In Fig. 10, we represented the optimal combination of 4, 6, 8, and 10 wavelengths for monitoring , , and changes in activated grey matter. These wavelengths are represented in Fig. 10(a) by dots. We also represented the configuration of four wavelengths proposed by Arifler et al.31 and the broadband spectral range used by Bale et al.14,17 The quantification errors are represented in Fig. 10(b) and the measurements of , , and are plotted in Figs. 10(c), 10(d), and 10(e), respectively. The minimum errors in , , and are obtained with the combination of 10 wavelengths calculated with our optimization procedure. With this digital instrument simulator, we modeled an exposed cortex. The simulated light is not absorbed by the skin or the skull of the patient, and the visible light can be collected by the camera. Thus, the optimization procedure can take profit of the peaks of oxidized and reduced CCO extinction spectra in the visible and the near-infrared range to monitor the three chromophores. The errors in , , and obtained with this configuration were 23%, 48%, and 23% lower than those obtained with broadband spectra between 780 and 900 nm, respectively. For all chromophores, the standard deviation of the RMSE decreases as more spectral bands are considered. For , the mean RMSE decreases slightly with the number of wavelengths, whereas the mean errors for and remain constant. For all wavelengths, the mean RMSE is lower than that computed with Bale’s spectral configuration. For and , the standard deviation is less than or equal to that obtained with Bale’s configuration. However, for , the standard deviation is higher than that obtained with Bale’s configuration. In Fig. 11, we represented the optimal combination of 6, 8, and 10 wavelengths for monitoring , , , , and changes in activated grey matter. These wavelengths are represented in Fig. 11(a) by dots. The quantification errors are represented in Fig. 11(b), and the concentration changes are plotted in Figs. 11(c), 11(d), 11(e), 11(f), and 11(g). The minimum errors to resolve all chromophores are obtained with the combination of 8 and 10 wavelengths. It is interesting to note that the use of a larger number of wavelengths (10) does not significantly reduce quantification errors. With the use of 6, 8, and 10 optimal wavelengths, the total quantification errors were 29, 28, and , respectively. The errors in and were higher than those obtained in Fig. 10. However, with the consideration of cytochromes and in the modified Beer–Lambert law, the errors in were 19% lower than those obtained in Fig. 10. The noise in the measurements of and is very important, making it difficult to differentiate between activated and non-activated grey matter. As proposed in our previous study,22 we can also identify different combinations of wavelengths to independently quantify the concentration changes in , Hb, oxCCO, oxCytb, and oxCytc, see Fig. 12. For instance, if we want to identify the optimal wavelength to resolve changes only, the optimization procedure can be run to minimize values by only considering three chromophores in the modified Beer–Lambert law [, Hb, and oxCytb, see Eq. (7)]. With this approach, we represented in Fig. 12(a) the quantification errors obtained with the optimal wavelength combination to separately resolve , Hb, oxCCO, oxCytb, and oxCytc changes. We also plotted the concentration changes in , Hb, oxCCO, oxCytb, and oxCytc in Figs. 12(b), 12(c), 12(d), 12(e), and 12(f). In Fig. S2 in the Supplementary Material, we represented the optimal combinations of 4, 6, 8, and 10 wavelengths to separately resolve , Hb, oxCCO, oxCytb, and oxCytc changes. The separate optimization procedure leads to different spectral configurations for each chromophore. For all chromophores (expect oxCytb), the mean RMSE decreases with the number of wavelengths. We obtained smaller quantification errors when we applied the optimization procedure on each chromophore separately than when we tried to resolve all the chromophores at the same time, see Table 3. Table 3Average of the quantification errors (RMSE in μMol) obtained with two Beer–Lambert systems proposed in the literature and optimal combination of six wavelengths.
4.DiscussionWe presented a digital instrument simulator to optimize the development of hyperspectral systems for intraoperative brain mapping studies. The code is publicly available on GitHub. This simulator is based on a realistic digital phantom of an exposed cortex computed with white Monte Carlo simulations. We developed C++ software to reconstruct temporal perturbations of the absorption coefficient using only one simulation. Using this Monte Carlo framework, we also proposed an optimization procedure based on the genetic algorithm to identify the best wavelength combinations in the visible and near-infrared range to quantify changes in , Hb, oxCCO, oxCytb, and oxCytc. This digital instrument simulator has been developed for intraoperative brain mapping studies. However, the code could be easily adapted for other studies. For an organ other than the brain, the segmentation step will certainly have to be adapted (segmentation method, number of classes), as well as the optical properties (see Table 1) and the chemical composition (see Table 2) for each class of the modeled tissue. 4.1.Digital Instrument SimulatorThe digital instrument allows the modeling of intensity maps collected by a camera sensor as well as the estimation of the mean path length of traveled photons through the tissue. This framework could be used to improve clinical and preclinical optical devices for brain mapping applications. In these studies, the modified Beer–Lambert law is used to monitor chromophore changes in an animal29,30 or a human7,9,11,52 brain. However, measurements are highly subject to quantification errors if incorrect path lengths are used to resolve chromophore changes. This could be the case in a lot of applications where the inhomogeneities of the optical properties are not taken into consideration. Indeed, a single path length is considered for the whole cerebral cortex, which is usually estimated with a homogeneous volume of grey matter with Monte Carlo simulations29 or using the analytical solution to the diffusion approximation of the radiative transfer equation in a semi-infinite geometry.30,52 In our study, we proposed to take into account the inhomogeneities of the optical properties with a pixel-wise estimation of the mean path length. As we can see on Figs. 7 and 8, the blood vessels have a great impact on the path length. For large blood vessels having a 2-mm diameter, the path length is almost null in the visible spectra (below 600 nm) and is lower than the path length estimated for grey matter in the red and near-infrared range. However, the path length measured at the level of the small blood vessels () is almost the same as that measured at the level of the grey matter. This means that the detected packets of photons are mostly propagated in the grey matter under these small blood vessels. This result is consistent with the results presented by Giannoni et al.24 and indicates that large blood vessels have a large impact on the path length. We also observed that there were no significant differences between the mean path length measured on arteries and veins. Indeed, when comparing the absorption coefficients of the artery and the vein (), we do not observe as great a difference as that between the artery and the grey matter () or the vein and the grey matter (, with the average absorption coefficient over the wavelength range 400 to 1000 nm). The segmentation step in the digital instrument simulator could also be improved, see Fig. 1 and Sec. 2.1.2. Segmentation is based on morphological operations calculated in a mask delimiting the surgical window. The latter was created manually. This manual operation has been chosen because automatic thresholding, using Otsu’s method53 for example, does not allow the cerebral cortex to be identified repeatedly. In some patients, bleeding around the edges of the surgical window or bloody compresses prevents the cerebral cortex from being correctly identified. To overcome this limitation, segmentation methods based on deep learning (U-Net)54 could be used for segmenting the blood vessels and delineate the contour of the surgical window. For the moment, this approach has not been implemented because we do not have enough data to train the neural network. In this study, we modeled a perfect optical device. The real spectral characteristics of the light source and the sensor could be integrated into the digital simulator by normalizing the estimated quantities with spectral sensitivity curves of the light sources and the camera, as proposed in our previous study.22 We modeled a real sensor in terms of resolution and definition, but this could be improved with the conversion of the diffuse reflectance into sensor intensities by considering parameters of the optics systems55 (lens magnification, lens transmission, quantum efficiency). We can also integrate the effect of the lenses on the image creation such as geometric distortion, vignetting, or chromatic aberrations. A lens design model can be used for this purpose.56 Using this digital instrument simulator, we modeled a simple cerebral activation by changing the concentrations of the chromophores in a portion of activated grey matter (see Fig. 5). However, we do not model the complex physiologic events related to the neuro-vascular coupling. To answer this limitation, we plan to model the spatio-temporal concentration changes in arteriole, capillary, and venous compartments related to blood volume, flow velocity, and oxygen consumption with the implementation of the dynamic model proposed by Fantini.23 Although the strength of this digital instrument is to use the white Monte Carlo approach, this method has also a limitation, which is the size of the generated files. Indeed, for one wavelength, 5 GB of data is generated, which leads to 305 GB for the 61 wavelengths modeled in our study. To reduce the memory size, the absorption coefficients need to be fixed in the modeled tissue to directly estimate the radiative quantities. However, with this approach, we cannot reconstruct the temporal changes of the absorption coefficient, so a simulation has to be performed per wavelength and per temporal index, which increases drastically the computation time. 4.2.Identification of the Optimal Wavelength for Hemodynamic and Metabolic MonitoringUsing the data obtained with the digital instrument simulator, we proposed an optimization procedure based on the genetic algorithm to identify the best wavelength combinations in the visible and near-infrared range to quantify changes in , Hb, oxCCO, oxCytb, and oxCytc. We added noise to the simulated quantities to identify the wavelength that is the most robust to noise. For hemodynamic monitoring, see Fig. 9, we can see that the optimal groups of 2, 4, 6, 8, or 10 wavelengths aim to reduce the quantification errors in and compared with that obtained with the configurations proposed in the literature. Contrary to the spectral configurations proposed by Bouchard et al.29 and White et al.,30 the wavelengths identified with the optimization procedure do not include the hemoglobin isobestic point at 530 nm. The wavelengths were located in the visible and near-infrared range, where or Hb absorption predominates. This could be explained by the addition of noise in the optimization procedure, which makes it difficult to interpret the attenuation changes measured at the isobestic points. Moreover, the optimization procedure identifies the best wavelengths but does not take into account the bandwidth of the light source spectra in hyperspectral devices based on spectral-scanning technology.28 Taking into account a spectral bandwidth could modify the value of the central wavelengths by a few nanometers but would not introduce changes greater than the spectral bandwidth. The optimization procedure took profit of the peaks of oxidized and reduced cytochrome extinction spectra in the visible and the near-infrared range to monitor hemodynamic and metabolic changes. For hemodynamics and oxCCO monitoring, the optimal wavelengths aim to reduce the quantification errors compared with that obtained with broadband spectra between 780 and 900 nm. In this study, we also proposed to quantify the changes of the cytochrome and . We can see in Fig. 11 that the addition of the cytochromes and in the modified Beer–Lambert system helps to better resolve oxCCO changes compared with the model based on three chromophores (, Hb, and oxCCO), see Fig. 10. Indeed, we can see that the quantification errors in oxCCO are lower when all cytochromes are considered, see Table 3. This result is interesting as it may help to obtain robust devices to monitor brain metabolism, which could be a hyperspectral device for intraoperative brain mapping or even a NIRS device for bedside monitoring. The monitoring of the oxidation state of CCO could help to obtain a more direct biomarker of neuronal activity, to detect brain injuries,16,57 and to better understand the neurovascular coupling.58 The optimization procedure does not help to resolve changes in cytochrome and in a significant way. Indeed, we can see in Figs. 11 and 12 that oxCytb and oxCytc measurements are really noisy. A two-sample -test tells us that changes measured at the level of the activated grey matter are not significantly different from those measured on non-activated grey matter ( are mainly higher than 0.1). This is maybe due to the noise addition, coupled with the fact that the attenuation changes due to the oxidized cytochrome and are rather low compared with those of hemoglobin and oxCCO. The optimization procedure helps to obtain smaller quantification errors compared with the literature configurations. For hemodynamic monitoring, increasing the number of wavelengths from 2 to 10 helps to reduce linearly the quantification error from 8% to 26%, see Fig. 9. However for hemodynamic and metabolic monitoring, we do not observe a significant decrease in the quantification error compared with Bale’s spectral configuration if using 6, 8, or 10 optimal wavelengths (quantification errors of 32%, 33%, and 33%, respectively, see Fig. 10). Thus, the number of wavelengths of a hyperspectral camera could be reduced to six to resolve hemodynamic and CCO changes. It could help to optimize several critical parameters for intraoperative functional brain mapping applications such as the signal-to-noise ratio. However, the availability of light sources or spectral filters at the identified wavelengths may limit the development of the hyperspectral camera. To address this limitation, we plan to restrict the optimization procedure to a number of wavelengths selected from the available light sources or spectral filters. For each wavelength, the spectral bandwidth can also be incorporated into the procedure. As proposed by Leadley et al,32 we will also investigate in a future study the influence of extinction coefficient uncertainties on chromophore quantification. We also showed that a separate quantification of hemodynamics and oxCCO leads to a better estimation of the chromophore changes, see Table 3. The optimization procedure helps to identify the best wavelength combination in the range 400 to 1000 nm, which reduces the quantification errors in , Hb, and oxCCO by 47%, 57%, and 57% compared with the gold standard of 121 wavelengths between 780 and 900 nm. The configuration of six wavelengths is 468, 482, 610, 753, 814, and 907 nm to monitor and Hb changes and 608, 650, 661, 672, 820, and 865 nm to monitor oxCCO changes. The separate quantification method should be taken with caution because it distorts the link between chromophores. Moreover, this could lead to incorrect interpretations between the chromophore changes and the physiological status of the patient. However, this method could be interesting for some clinical applications. For example, the method can be used to identify a spectral configuration for precise monitoring of Hb changes with the idea to define a blood oxygen level dependent-like contrast.58 In intraoperative brain mapping studies, this approach has been proposed by several research groups. The authors used a single illumination at 60559 or 610 nm60 combined with a monochrome camera to assess Hb absorption. For these two wavelengths, the authors considered that the signal changes were mainly due to Hb because the absorption related to is negligible compared with that of Hb (the ratio between Hb molar extinction coefficient to that of is ). Although the contribution of is small compared with that of Hb, the signal measured with this type of device cannot accurately measure variations in Hb concentration. Our method solves this problem. 5.ConclusionIn this paper, we proposed two main contributions. First, we presented a digital instrument simulator to optimize the development of hyperspectral systems for intraoperative brain mapping studies. This simulator is based on a realistic digital phantom of an exposed cortex computed with white Monte Carlo simulations. A C++ software was developed to reconstruct temporal perturbations of the absorption coefficient using only one simulation. Second, we presented an optimization procedure based on the genetic algorithm to identify the best wavelength combinations in the visible and near-infrared range to quantify changes in , Hb, oxCCO, oxCytb, and oxCytc. With the digital instrument simulator, we can improve the accuracy of actual preclinical and clinical optical devices used in brain mapping applications with the consideration of the impact of the large blood vessels on the path length. We also proposed several spectral configurations to monitor hemodynamic and metabolic changes in grey matter that aimed to reduce the quantification error of changes in , Hb, and oxCCO compared with the configuration proposed in the literature. This digital instrument simulator and this optimization framework could be used to optimize the design of hyperspectral imaging devices,61,62 with the objective of transforming current practice toward an all-optical, real-time, quantitative, and accurate imaging approach, which could significantly help neurosurgeons, enhance the efficacy of the treatment, and ultimately improve the quality of life and functional outcomes of the patients after the brain tumor resection. Code and Data AvailabilityIn support of open science, the code presented in this article is publicly available on https://github.com/CCaredda/White-Monte-Carlo/. AcknowledgmentsThese works were funded by the European Union’s Horizon Europe Research and Innovation Programme (Grant No. 101071040)—project HyperProbe; LABEX PRIMES (Grant No. ANR-11-LABX-0063) of Université de Lyon, within the program “Investissements d’Avenir” (Grant No. ANR-11-IDEX-0007), operated by the French National Research Agency (ANR); Infrastructures d’Avenir en Biologie Santé (Grant No. ANR-11-INBS-000), within the program “Investissements d’Avenir” operated by the French National Research Agency (ANR) and France Life Imaging (Grant No. ANR-11-INBS-0006). FL and IT are supported by UCL, which, as a UK participant in Horizon Europe Project HyperProbe is supported by UKRI (Grant No. 10048387). We want to acknowledge the PILoT facility for the support provided for the image acquisition. This material is based upon work done on the ISO 9001:2015 PILoT facility. ReferencesS. Ogawa et al.,
“Brain magnetic resonance imaging with contrast dependent on blood oxygenation,”
Proc. Natl. Acad. Sci., 87 9868
–9872 https://doi.org/10.1073/pnas.87.24.9868
(1990).
Google Scholar
I. J. Gerard et al.,
“Brain shift in neuronavigation of brain tumors: a review,”
Med. Image Anal., 35 403
–420 https://doi.org/10.1016/j.media.2016.08.007
(2017).
Google Scholar
T. Gasser et al.,
“Intraoperative functional MRI: implementation and preliminary experience,”
NeuroImage, 26 685
–693 https://doi.org/10.1016/j.neuroimage.2005.02.022 NEIMEF 1053-8119
(2005).
Google Scholar
C. Schulz, S. Waldeck and U. M. Mauer,
“Intraoperative image guidance in neurosurgery: development, current indications, and future trends,”
Radiol. Res. Pract., 2012
(1), 197364 https://doi.org/10.1155/2012/197364
(2012).
Google Scholar
E. Mandonnet, P. A. Winkler and H. Duffau,
“Direct electrical stimulation as an input gate into brain functional networks: principles, advantages and limitations,”
Acta Neurochirurg., 152 185
–193 https://doi.org/10.1007/s00701-009-0469-0
(2010).
Google Scholar
J. Pallud et al.,
“Technical principles of direct bipolar electrostimulation for cortical and subcortical mapping in awake craniotomy,”
Neurochirurgie, 63 158
–163 https://doi.org/10.1016/j.neuchi.2016.12.004
(2017).
Google Scholar
C. Caredda et al.,
“Intraoperative identification of functional brain areas with RGB imaging using statistical parametric mapping: simulation and clinical studies,”
NeuroImage, 278 120286 https://doi.org/10.1016/j.neuroimage.2023.120286 NEIMEF 1053-8119
(2023).
Google Scholar
B. Chance et al.,
“Cognition-activated low-frequency modulation of light absorption in human brain,”
Proc. Natl. Acad. Sci., 90 3770
–3774 https://doi.org/10.1073/pnas.90.8.3770
(1993).
Google Scholar
C. Caredda et al.,
“Intraoperative quantitative functional brain mapping using an RGB camera,”
Neurophotonics, 6
(4), 045015 https://doi.org/10.1117/1.NPh.6.4.045015
(2019).
Google Scholar
C. Caredda et al.,
“Real time intraoperative functional brain mapping based on RGB imaging,”
IRBM, 42
(3), 189
–197 https://doi.org/10.1016/j.irbm.2020.04.004
(2021).
Google Scholar
C. Caredda et al.,
“Intraoperative resting-state functional connectivity based on RGB imaging,”
Diagnostics, 11
(11), 2067 https://doi.org/10.3390/diagnostics11112067
(2021).
Google Scholar
C. Caredda et al.,
“Intraoperative functional and metabolic brain mapping using hyperspectral imaging,”
Proc. SPIE, 11225 112250B https://doi.org/10.1117/12.2545968 PSISDG 0277-786X
(2020).
Google Scholar
L. Giannoni et al.,
“A hyperspectral imaging system for mapping haemoglobin and cytochrome-c-oxidase concentration changes in the exposed cerebral cortex,”
IEEE J. Sel. Top. Quantum Electron., 27
(4), 1
–11 https://doi.org/10.1109/JSTQE.2021.3053634 IJSQEN 1077-260X
(2021).
Google Scholar
G. Bale, C. E. Elwell and I. Tachtsidis,
“From Jöbsis to the present day: a review of clinical near-infrared spectroscopy measurements of cerebral cytochrome-c-oxidase,”
J. Biomed. Opt., 21
(9), 091307 https://doi.org/10.1117/1.JBO.21.9.091307 JBOPFO 1083-3668
(2016).
Google Scholar
B. Kadenbach,
“Regulation of mitochondrial respiration and ATP synthesis via cytochrome c oxidase,”
Rendiconti Lincei. Scienze Fisiche e Naturali, 29 421
–435 https://doi.org/10.1007/s12210-018-0710-y
(2018).
Google Scholar
G. Bale, S. Mitra and I. Tachtsidis,
“Metabolic brain measurements in the newborn: advances in optical technologies,”
Physiol. Rep., 8
(17), e14548 https://doi.org/10.14814/phy2.14548
(2020).
Google Scholar
G. Bale et al.,
“A new broadband near-infrared spectroscopy system for in-vivo measurements of cerebral cytochrome-c-oxidase changes in neonatal brain injury,”
Biomed. Opt. Express, 5
(10), 3450
–3466 https://doi.org/10.1364/BOE.5.003450 BOEICL 2156-7085
(2014).
Google Scholar
B. W. Pogue and M. S. Patterson,
“Review of tissue simulating phantoms for optical spectroscopy, imaging and dosimetry,”
J. Biomed. Opt., 11
(4), 041102 https://doi.org/10.1117/1.2335429 JBOPFO 1083-3668
(2006).
Google Scholar
S. Kleiser et al.,
“Comparison of tissue oximeters on a liquid phantom with adjustable optical properties: an extension,”
Biomed. Opt. Express, 9 86
–101 https://doi.org/10.1364/BOE.9.000086 BOEICL 2156-7085
(2018).
Google Scholar
F. Lange et al.,
“Maestros: a multiwavelength time-domain NIRs system to monitor changes in oxygenation and oxidation state of cytochrome-c-oxidase,”
IEEE J. Sel. Top. Quantum Electron., 25
(1), 1
–12 https://doi.org/10.1109/JSTQE.2018.2833205 IJSQEN 1077-260X
(2018).
Google Scholar
A. Sudakou et al.,
“Time-domain NIRs system based on supercontinuum light source and multi-wavelength detection: validation for tissue oxygenation studies,”
Biomed. Opt. Express, 12 6629
–6650 https://doi.org/10.1364/BOE.431301 BOEICL 2156-7085
(2021).
Google Scholar
C. Caredda et al.,
“Optimal spectral combination of a hyperspectral camera for intraoperative hemodynamic and metabolic brain mapping,”
Appl. Sci., 10
(15), 5158 https://doi.org/10.3390/app10155158
(2020).
Google Scholar
S. Fantini,
“Dynamic model for the tissue concentration and oxygen saturation of hemoglobin in relation to blood volume, flow velocity, and oxygen consumption: implications for functional neuroimaging and coherent hemodynamics spectroscopy (CHS),”
NeuroImage, 85 202
–221 https://doi.org/10.1016/j.neuroimage.2013.03.065 NEIMEF 1053-8119
(2014).
Google Scholar
L. Giannoni, F. Lange and I. Tachtsidis,
“Investigation of the quantification of hemoglobin and cytochrome-c-oxidase in the exposed cortex with near-infrared hyperspectral imaging: a simulation study,”
J. Biomed. Opt., 25
(4), 046001 https://doi.org/10.1117/1.JBO.25.4.046001 JBOPFO 1083-3668
(2020).
Google Scholar
D. A. Boas et al.,
“The accuracy of near infrared spectroscopy and imaging during focal changes in cerebral hemodynamics,”
NeuroImage, 13
(1), 76
–90 https://doi.org/10.1006/nimg.2000.0674 NEIMEF 1053-8119
(2001).
Google Scholar
S. Matcher et al.,
“Performance comparison of several published tissue near-infrared spectroscopy algorithms,”
Anal. Biochem., 227
(1), 54
–68 https://doi.org/10.1006/abio.1995.1252 ANBCA2 0003-2697
(1995).
Google Scholar
B. Fei,
“Hyperspectral imaging in medical applications,”
Data Handling in Science and Technology, 32 523
–565 Elsevier(
(2019). Google Scholar
G. Lu and B. Fei,
“Medical hyperspectral imaging: a review,”
J. Biomed. Opt., 19
(1), 010901 https://doi.org/10.1117/1.JBO.19.1.010901 JBOPFO 1083-3668
(2014).
Google Scholar
M. B. Bouchard et al.,
“Ultra-fast multispectral optical imaging of cortical oxygenation, blood flow, and intracellular calcium dynamics,”
Opt. Express, 17 15670
–15678 https://doi.org/10.1364/OE.17.015670 OPEXFF 1094-4087
(2009).
Google Scholar
B. R. White et al.,
“Statistical approaches to temporal and spatial autocorrelation in resting-state functional connectivity in mice measured with optical intrinsic signal imaging,”
Neurophotonics, 9
(4), 041405 https://doi.org/10.1117/1.NPh.9.4.041405
(2022).
Google Scholar
D. Arifler et al.,
“Optimal wavelength combinations for near-infrared spectroscopic monitoring of changes in brain tissue hemoglobin and cytochrome c oxidase concentrations,”
Biomed. Opt. Express, 6
(3), 933
–947 https://doi.org/10.1364/BOE.6.000933 BOEICL 2156-7085
(2015).
Google Scholar
G. Leadley et al.,
“Investigating the effect of limited spectral information on NIRS-derived changes in hemoglobin and cytochrome-c-oxidase concentration with a diffusion-based model,”
Biomed. Opt. Express, 15
(10), 5912
–5931 https://doi.org/10.1364/BOE.531775 BOEICL 2156-7085
(2024).
Google Scholar
G. Bradski,
“The OpenCV library,”
(2000). https://github.com/opencv/opencv/wiki/CiteOpenCV Google Scholar
Q. Fang and D. A. Boas,
“Monte Carlo simulation of photon migration in 3D turbid media accelerated by graphics processing units,”
Opt. Express, 17
(22), 20178
–20190 https://doi.org/10.1364/OE.17.020178 OPEXFF 1094-4087
(2009).
Google Scholar
S. L. Jacques,
“Optical properties of biological tissues: a review,”
Phys. Med. Biol., 58
(11), R37 https://doi.org/10.1088/0031-9155/58/11/R37 PHMBA7 0031-9155
(2013).
Google Scholar
A. Yaroslavsky et al.,
“Optical properties of selected native and coagulated human brain tissues in vitro in the visible and near infrared spectral range,”
Phys. Med. Biol., 47
(12), 2059 https://doi.org/10.1088/0031-9155/47/12/305 PHMBA7 0031-9155
(2002).
Google Scholar
D. Sardar and L. Levy,
“Optical properties of whole blood,”
Lasers Med. Sci., 13 106
–111 https://doi.org/10.1007/s101030050062
(1998).
Google Scholar
J. Binding et al.,
“Brain refractive index measured in vivo with high-NA defocus-corrected full-field OCT and consequences for two-photon microscopy,”
Opt. Express, 19
(6), 4833
–4847 https://doi.org/10.1364/OE.19.004833 OPEXFF 1094-4087
(2011).
Google Scholar
A. Sassaroli and F. Martelli,
“Equivalence of four Monte Carlo methods for photon migration in turbid media,”
J. Opt. Soc. Am. A, 29
(10), 2110
–2117 https://doi.org/10.1364/JOSAA.29.002110 JOAOD6 0740-3232
(2012).
Google Scholar
T. Correia, A. Gibson and J. Hebden,
“Identification of the optimal wavelengths for optical topography: a photon measurement density function analysis,”
J. Biomed. Opt., 15
(5), 056002 https://doi.org/10.1117/1.3484747 JBOPFO 1083-3668
(2010).
Google Scholar
H. Mitchell et al.,
“The chemical composition of the adult human body and its bearing on the biochemistry of growth,”
J. Biol. Chem., 158
(3), 625
–637 https://doi.org/10.1016/S0021-9258(19)51339-4 JBCHA3 0021-9258
(1945).
Google Scholar
L. Gagnon et al.,
“Double-layer estimation of intra-and extracerebral hemoglobin concentration with a time-resolved system,”
J. Biomed. Opt., 13
(5), 054019 https://doi.org/10.1117/1.2982524 JBOPFO 1083-3668
(2008).
Google Scholar
N. M. Wiseman et al.,
“Chapter 18 - perfusion and susceptibility weighted imaging in traumatic brain injury,”
Neurosensory Disorders in Mild Traumatic Brain Injury, 303
–319 Academic Press(
(2019). Google Scholar
T.-C. Poon and T. Kim, Engineering Optics with Matlab, World Scientific(
(2006). Google Scholar
D. A. Boas et al.,
“Three dimensional Monte Carlo code for photon migration through complex heterogeneous media including the adult human head,”
Opt. Express, 10 159
–170 https://doi.org/10.1364/OE.10.000159 OPEXFF 1094-4087
(2002).
Google Scholar
R. Yao, X. Intes and Q. Fang,
“Direct approach to compute Jacobians for diffuse optical tomography using perturbation Monte Carlo-based photon,”
Biomed. Opt. Express, 9 4588
–4603 https://doi.org/10.1364/BOE.9.004588 BOEICL 2156-7085
(2018).
Google Scholar
Y. Yuan et al.,
“Graphics processing units-accelerated adaptive nonlocal means filter for denoising three-dimensional Monte Carlo photon transport simulations,”
J. Biomed. Opt., 23
(12), 121618 https://doi.org/10.1117/1.JBO.23.12.121618 JBOPFO 1083-3668
(2018).
Google Scholar
Y. Ma et al.,
“Wide-field optical mapping of neural activity and brain haemodynamics: considerations and novel approaches,”
Philos. Trans. R. Soc. B: Biol. Sci., 371
(1705), 20150360 https://doi.org/10.1098/rstb.2015.0360 PTRBAE 0962-8436
(2016).
Google Scholar
P. Phan et al.,
“Spatial distribution of changes in oxidised cytochrome c oxidase during visual stimulation using broadband near infrared spectroscopy imaging,”
Oxygen Transport to Tissue XXXVIII, 195
–201 Springer(
(2016). Google Scholar
M. Mitchell, An Introduction to Genetic Algorithms, MIT Press(
(1998). Google Scholar
P. Virtanen et al.,
“SciPy 1.0: fundamental algorithms for scientific computing in Python,”
Nat. Methods, 17 261
–272 https://doi.org/10.1038/s41592-019-0686-2 1548-7091
(2020).
Google Scholar
J. Pichette et al.,
“Intraoperative video-rate hemodynamic response assessment in human cortex using snapshot hyperspectral optical imaging,”
Neurophotonics, 3
(4), 045003 https://doi.org/10.1117/1.NPh.3.4.045003
(2016).
Google Scholar
N. Otsu,
“A threshold selection method from gray-level histograms,”
IEEE Trans. Syst. Man Cybern., 9
(1), 62
–66 https://doi.org/10.1109/TSMC.1979.4310076
(1979).
Google Scholar
A. O. Joshua, F. V. Nelwamondo and G. Mabuza-Hocquet,
“Blood vessel segmentation from fundus images using modified U-Net convolutional neural network,”
J. Image Graph., 8
(1), 21
–25 https://doi.org/10.18178/joig.8.1.21-25
(2020).
Google Scholar
J. E. Farrell, P. B. Catrysse and B. A. Wandell,
“Digital camera simulation,”
Appl. Opt., 51
(4), A80
–A90 https://doi.org/10.1364/AO.51.000A80 APOPAI 0003-6935
(2012).
Google Scholar
P. Y. Maeda, P. B. Catrysse and B. A. Wandell,
“Integrating lens design with digital camera simulation,”
Proc. SPIE, 5678 48
–58 https://doi.org/10.1117/12.588153 PSISDG 0277-786X
(2005).
Google Scholar
A. Bainbridge et al.,
“Brain mitochondrial oxidative metabolism during and after cerebral hypoxia-ischemia studied by simultaneous phosphorus magnetic-resonance and broadband near-infrared spectroscopy,”
NeuroImage, 102 173
–183 https://doi.org/10.1016/j.neuroimage.2013.08.016 NEIMEF 1053-8119
(2013).
Google Scholar
E. M. Hillman,
“Coupling mechanism and significance of the bold signal: a status report,”
Annu. Rev. Neurosci., 37 161
–181 https://doi.org/10.1146/annurev-neuro-071013-014111 ARNSD5 0147-006X
(2014).
Google Scholar
K. Sato et al.,
“Intraoperative intrinsic optical imaging of human somatosensory cortex during neurosurgical operations,”
Neurophotonics, 4
(3), 031205 https://doi.org/10.1117/1.NPh.4.3.031205
(2017).
Google Scholar
A. F. Cannestra et al.,
“Temporal and topographical characterization of language cortices using intraoperative optical intrinsic signals,”
NeuroImage, 12
(1), 41
–54 https://doi.org/10.1006/nimg.2000.0597 NEIMEF 1053-8119
(2000).
Google Scholar
|