Open Access
4 June 2024 Photothermal gas sensing with a custom-made, fiber-coupled Fabry–Pérot etalon
Author Affiliations +
Abstract

Growing awareness of the adverse health effects of air pollution has increased the demand for reliable, sensitive, and mass-producible sensor systems. Photothermal interferometry has shown great promise for sensitive, selective, and miniaturized gas sensing solutions. This work describes the development of a macroscopic photothermal sensor system with a sensor head consisting of a low-cost, custom-made, and fiber-coupled Fabry–Pérot etalon. The sensor was tested with NO2, achieving a 3σ limit of detection (LOD) of approximately 370 ppbv (1 s). Exhibiting little drift, a LOD of 15 ppbv is achievable for 200 s integration time. Compensating for the excitation power, the normalized noise equivalent absorption was calculated to be 1.4×108 cm1WHz. The sensor system is not limited to NO2 but can be used for any gas or aerosol species by exchanging the excitation laser source.

1.

Introduction

The detection of trace concentrations of gases is crucial in countless fields of application. From environmental monitoring to leakage detection in industrial applications or in the medical sector, sensitive and selective determination of gas concentrations has to be ensured in a cost-effective way. There are many known gas sensing methods, but they either lack the necessary accuracy or do not meet the price requirements for the respective application.1 Laser-based spectroscopic sensor systems offer the advantage of a high sensitivity and selectivity.2 In tunable diode laser based absorption spectroscopy, the signal scales with the gas-laser beam interaction length according to the Beer–Lambert law. It states that the incident laser power is attenuated exponentially over the detection path length, thereby posing a limit to the miniaturization potential.3 In photothermal spectroscopy on the other hand, the signal detection can be done with interferometric precision and sensitivity, enabling the realization of just minute detection volumes.48 However, most of the photothermal interferometry (PTI) setups presented to date, are large and rather complex bench-top systems relying on the use of expensive equipment. For field application, the sensing system needs to be down-scaled and a rugged interferometric sensing unit is essential. For the characterization of the setup proposed in this work, we have chosen nitrogen dioxide (NO2) as analyte, whose detrimental effect on both health and climate has been proven in many studies.9 Evidence suggests that elevated exposure levels can be associated with respiratory diseases as well as premature deaths.10 Precise monitoring of gas concentration levels is essential for raising public and political awareness to the problem of air pollution, and thus the enforcement of strict regulation standards. Optical sensors, such as photothermal and photoacoustic gas sensors, allow selective measurements and sensitivities down to the parts per billion by volume (ppbv) region have already been demonstrated.1115 This work presents the characterization of a novel optical fiber-based PTI gas sensing system. Instead of relying on a commercially available and expensive Fabry–Pérot etalon (FPE), we make use of a low-cost and custom-made FPE. The etalon has been manufactured with an easily accessible fabrication method using only basic photonic equipment, available in any photonic laboratory.16 The price of commercially available high-finesse FPEs surmounts the commonly acceptable budget for prototyping. By relying on off-the-shelf bulk-optic and telecom fiber-optic components, the material costs of the FPE, as used here, could be reduced to less than 400 euros. A comparison to similar configurations is difficult, given the lack of disclosed details about the fabrication methods as well as the equipment necessary for glass processing and coating. See, e.g., reference Chen et al.17 Convenient handling of the main sensing unit is ensured by fiber-coupling the air-spaced etalon. The presented gas sensing system is calibrated with well-defined concentrations of NO2, demonstrating its potential for various gas sensing applications. The calibration results also more generally display the favorable capabilities of the developed low-cost FPEs as a required next step toward low-cost sensitive photothermal trace gas monitoring.

2.

Methods

2.1.

PTI Working Principle

In PTI, minute changes in refractive index (RI), induced by exciting and thereby heating the target gas periodically via laser irradiation, are measured interferometrically.18 Careful selection of the excitation wavelength with respect to cross-sensitivities to other gases ensures a high selectivity toward the target gas while being unaffected by other gases, which may be present during the measurement. Periodic modulation of the excitation laser, either in wavelength or in intensity, leads to modulated absorption and RI changes, proportional to the gas concentration present. Neglecting excitational relaxation losses, the oscillatory temperature ΔTosc and phase Δϕosc amplitudes, induced in the gas by intensity-modulated and collimated laser irradiation, can be calculated with the following equations:18

Eq. (1)

ΔTosc=2αPeffπw2ρ0cpω,Δnosc=n01T0ΔTosc,
where α is the absorption coefficient of the gas, Peff the effective optical power of the excitation laser (peak-to-peak power), w the excitation laser’s beam radius, ρ0 and cp the density and specific heat capacity of air, and ω is the angular excitation frequency. These changes in RI are then detected, e.g., using an FPE, as illustrated in Fig. 1. In order to perform wavelength modulation spectroscopy (WMS), distinct absorption peaks are beneficial.19 In the region around 450 nm, NO2 exhibits strong but broad absorption features20 not suitable for WMS, which is why intensity modulation was performed for the measurements presented here.

Fig. 1

Exemplary reflection function of an FPE, plotted over the phase ϕ=(4πnd/λ). The periodic change of RI leads to a periodic change of the phase inside the etalon and therefore a modulation of the backreflected light ΔR. Given that the slope of the function can be assumed linear (dotted line) around the inflection point (black dot), ΔR is proportional to the gas concentration NNO2.

OE_63_6_064101_f001.png

The transmittance T and reflectance R=1T through the FPE can be described by the Airy distribution.21 Depending on the reflectivity Rm[0,1] of the identical mirrors and the phase ϕ, the transmittance for the case of normal incidence can be written as

Eq. (2)

T=(1Rm)212Rmcos(ϕ)+Rm2,ϕ=4πndλ,
where n is the RI, d is the distance between the two mirrors, and λ is the wavelength of the etalon’s probe laser. The red dotted line in Fig. 1 depicts the gradient of the function, here referred to as sensitivity S=(R/ϕ), which acts as conversion factor between the generated phase change and the subsequent change in reflected intensity. For photothermal measurements, the operation point of the interferometric detection unit is chosen to be at the inflection point of the function, where the sensitivity to phase changes is the highest. The sensitivity is proportional to the finesse of the FPE, which is commonly used as a figure of merit of an FPE. The use of such multi-pass interferometers allows miniaturization of the detection unit of a PTI sensor, by reducing the necessary gas-laser interaction path. By proper selection of the wavelength of the excitation laser, minimum cross-interferences with other gases, in terms of absorption as well as excitational relaxation pathways can be ensured. This makes PTI gas sensors highly selective and the interchangeable nature of the excitation laser allows detection of many gas and aerosol species, given the right laser source is available.

2.2.

Experimental Methods

A schematic of the macroscopic sensor system is shown in Fig. 2. The fiber-optic interferometric sensing part is depicted in the lower left corner of the figure. To minimize the influence of laser phase noise on the setup, an ultra-narrow linewidth fiber-coupled laser source (NKT, E15, 1550 nm) is used as probe laser. An optical isolator (AFW Technologies, SM dual-stage isolator) is installed, to prevent back-reflections into the laser. As a balanced detection (BD) scheme is used, the light intensity is subsequently split by a 90:10 fiber-optic splitter (Agiltron, SM micro-optic coupler) into a reference and signal path. Roughly 10 % of the light is directed onto the reference photodiode of the balanced detector (Newport, Nirvana 2017), whereas the rest passes a circulator (AFW Technologies, SM circulator) before entering the FPE.

Fig. 2

A schematic of the developed macroscopic sensor system visualizing its working principle. In the center, the 3D-printed measurement cell with the integrated fiber-coupled FPE is depicted.

OE_63_6_064101_f002.png

Etalons act as converter from frequency modulation, due to the laser’s phase noise, to amplitude modulation (FM to AM noise). As the FPE used for detection of the photothermal signal, is only present in the signal path, BD cannot compensate for the additional amplitude noise, highlighting the need for an ultra-low-noise laser source. Alternatively, a feedback-based stabilization scheme could be implemented, e.g., Pound–Drever–Hall method, with the advantage of having a larger tolerance toward low-cost laser sources.22 However, additional optical and electronical components would be required, adding to the complexity and size of the setup. To allow interaction between the target gas and the laser light, an air-spaced FPE has to be used. As commercially available FPEs are typically expensive, a novel low-cost method to fabricate fiber-coupled, air-spaced FPEs was developed, to allow fast and easy prototyping. Details of the fabrication process can be found in Tanzer et al.16 The FPE used for the NO2 measurements, as presented here, has a finesse of roughly 30 and a sensitivity to RI changes of 8.2·106%RIU. Direct mounting of the FPE to a single-mode fiber collimator (DK photonics, SM fiber collimator) creates a high mechanical stability of the detection unit, without the need for re-alignment after handling of the measurement cell. A picture of the measurement cell is shown in Fig. 3.

Fig. 3

Picture of the measurement cell with the excitation laser turned on.

OE_63_6_064101_f003.png

A gradient index lens collimates the laser beam out of the fiber and collects the reflected light from the FPE. The circulator then directs the reflected light onto the signal photodiode of the balanced detector. The excitation laser (Laser Components, PLT5 450B, 450 nm, 80 mW) beam is focused perpendicularly onto the probe laser beam inside the air-spaced etalon. Two wedged windows ensure that the measurement cell is air-tight while allowing the excitation beam to pass. For data acquisition and management of the system, an FPGA-based system (NI, cRIO9031) with a 24 bit ADC (NI, 9234) is used. This system controls the laser drivers, acquires data from the detectors and also features a lock-in amplifier (LIA), which allows for phase sensitive background correction of the photothermal signal amplitude. Operation of the system is done via a custom LabView-based user interface. To maintain the probe laser’s wavelength at the FPE transfer function’s inflection point and ensure optimum sensitivity of the system during measurements, a PID-controller was implemented. A fast response of the controller has to be ensured, to avoid drift of the operation point when the gas concentration or composition changes throughout the measurement campaign. For the characterization of the instrument with well-defined NO2 concentrations, a custom-made gas diluter, based on critical orifices, is used, to dilute 1000 ppmv of NO2 with pure nitrogen N2 down to the desired concentration levels.23 As abrupt changes of the flow rate can lead to a strong drift of the operation point, a pump in combination with a critical orifice (0.249 slm) is installed downstream of the cell. Together with a venting tube upstream of the cell, constant flow through the FPE is ensured. For environmental sensing applications however, fast changes of pressure and flow rate through the system can be avoided, making the use of these additional components obsolete. To achieve maximum sensitivity, the operation parameters of the setup were determined in the following way: with the maximum NO2 concentration present in the measurement cell, a frequency sweep of the intensity modulated excitation laser (50% duty cycle) was performed, to determine the optimal excitation frequency. As longer integration times lower the limit of detection (LOD), stability measurements were performed for more than 1 h. By calculating the Allan deviation, the maximum integration time was determined.24,25

3.

Results

By comparing the measured signal-to-noise ratios (SNRs) obtained via the frequency sweep, the ideal excitation frequency fexc was determined, as depicted in Fig. 4. The fluctuations in the measured data points can be attributed to the limited integration time of the LIA of 5 s. Repeated measurements for the regions of interest around 650 and 1100 nm with an integration time of 10 s revealed an optimum excitation frequency of 1065 Hz.

Fig. 4

The SNRs as obtained by sweeping the excitation frequency from 100 to 2000 Hz during measurements of 1000 ppmv of NO2 with 5 s integration time.

OE_63_6_064101_f004.png

In Fig. 5, the measured background-corrected lock-in amplitude for an integration time τint of 1 s is shown over the NO2 concentration as set by the gas diluter.

Fig. 5

The measured LIA amplitude for different NO2 concentrations with 1 s integration time. The black line is a linear fit and indicates the linearity of the sensor system. As only one measurement was performed per data point, no standard deviation can be given. The uncertainties of the gas concentrations are too small to be visible in the plot.

OE_63_6_064101_f005.png

The black line is a first order polynomial fit with zero offset, indicating the linearity of the sensor system. The inset on the upper left side of the plot shows the low concentration region up to 60 ppmv. As only one measurement per data point was performed, no statistical evaluation can be given.

The main source of uncertainty in the measurement results can be attributed to the reference system, namely the gas diluter and the subsequent tubing system for supplying gas to the measurement cell. The uncertainties of the concentrations, as given by the gas diluter, are plotted as errorbars in x-direction, however, they are too small to be visible in the graph. The deviations from the linear fit, especially in the low concentration regions, are mainly stemming from insufficient gas exchange times. The long tubing and possible dead volumes in the measurement cell demand flushing times for multiple hours, which was not practical during the measurement campaign. As no reference sensor was used, no quantification thereof can be given at this point. In Table 1 a list of sources adding to the uncertainty of the measurement is given, together with a ranking in terms of relative weight of the respective source. For most of the indicated sources, mitigation measures have been taken. To address fluctuations of the sensitivity of the FPE, a PID-controller based operation point stabilization was implemented, as mentioned before. To avoid excitation laser power fluctuations, low-noise drivers were employed. Mechanical instabilities causing a changing overlap of probe and excitation laser were avoided by sturdy mounting. From Eq. 1, the theoretical influence of both last-mentioned factors can be assessed (ΔPeff and Δw). Parasitic etalon effects stemming from reflections of fiber connectors were reduced by the use of index matching gel. Non-ideal behavior of the fiber-optic components (isolator, splitter, and circulator) has to be considered and they have to be chosen according to the highest possible standards. The use of an ultra-low-noise laser source in combination with a BD scheme reduces the influence of probe laser-related noise on the measurement. Finally, as only low-noise electronics were used, the relative weight of electronics noise is ranked as minute.

Table 1

Main sources of uncertainty of the measurement results.

Sources of uncertaintyBrief descriptionWeight
GasGas bottle concentration997 ppmv of NO2 with 2 % uncertainty****
Gas diluterUncertainty of mixing ratio (0.1% to 1.2%)****
Incomplete gas exchangeLong required flushing times of tubing and dead volumes, especially for low concentrations****
SystemSensitivity fluctuationsVariable conditions (p,T,n) can change the slope of the FPE**
Exc. laser power instabilityFluctuating optical power emitted by exc. laser**
Mechanical instabilitiesVariable optical alignment can change overlap between exc. and probe laser*
Fiber-optical componentsUnwanted interference effects changing the reflected optical power*
Probe laser noisePhase noise, relative intensity noise, etc.*
Electronics noiseLaser drivers, detectors, data aquisition, etc.*

With the measured SNRs for each individual concentration, a mean 3σ LOD of roughly 370 ppbv (1 s integration time) could be achieved. To allow for comparison to previously reported setups, the normalized noise equivalent absorption (NNEA) is calculated, independent of the optical laser power and bandwidth. For the effective optical power Peff of 40 mW and 1 s integration time, we achieve an NNEA=αPeffSNRΔf of 1.4×108  cm1WHz.26 Here, α is the absorption coefficient of the gas and the measurement noise bandwidth Δf was approximated by a first-order filter noise bandwidth of Δf=πfc2=14τint for the LIA’s low-pass filter.27 To investigate the stability of the sensor system, measurements of the background signal, with a constant flow of N2, were performed for more than 1 hour. By calculating the Allan deviation and converting it into units of NO2 concentration, as shown in Fig. 6, the LOD for longer integration times can be evaluated.

Fig. 6

The black line indicates the calculated Allan deviation in units of NO2 concentration as measured from the background signal. The gray dotted lines depict the standard error.

OE_63_6_064101_f006.png

The system shows good stability up to 200 s, until drift starts to dominate. With an integration time of 200 s a theoretical 3σ LOD of approximately 15 ppbv can be achieved. Similar long-term measurements of the photothermal signal with NO2 present revealed stronger instabilities. Ongoing work indicates that additional, parasitic etalon effects, stemming from fiber-optic components and connectors, on top of the measurement FPE’s transfer function, lead to varying sensitivity. Additional factors influencing the stability during a photothermal measurement are the fluctuations and drift of the excitation laser’s optical power and mechanical instabilities of the setup, leading to a degradation of the overlap between probe and excitation laser.

4.

Conclusions and Outlook

In this paper, an interferometric gas sensing system utilizing the photothermal effect has been presented. As a detection unit, a custom-made fiber-coupled FPE was employed, which is based on a fabrication method for low-cost fiber-coupled FPEs.16 By measuring the reflected light of the FPE, one single optical fiber is sufficient to lead light in and out of the etalon, allowing easier handling of the sensor head and simplifying the alignment process. The system was characterized with NO2 concentrations ranging from 1 to 1000 ppmv. With 1 s integration time, a LOD of approximately 370 ppbv can be achieved. The stability of the sensor system allows for integration times up to 200 s, thereby lowering the LOD down to approximately 15 ppbv. To benchmark the system against previously published results, the NNEA was calculated to be 1.4×108  cm1WHz, thereby proving to be competitive, despite the used low-cost FPE. An additional advantage of the proposed sensing system is the interchangeable nature of the excitation laser source. Basically any absorbing gas or aerosol species can be detected, given the appropriate excitation laser source is available. By introducing multiple excitation beams with different wavelengths, the system could also be easily adapted for multi-gas detection. Observed instabilities of the photothermal signal are related to parasitic etalon effects, introduced by reflecting surfaces of connectors and other optical components. These effects are added on top of the FPE’s transfer function, leading to signal non-linearities. Further investigations to quantify the influence of said effect are still ongoing. The use of index-matching gel reduced the effect significantly and splicing of the fiber connections is expected to suppress unwanted etalon effects below the level of detection. Effort is also put into the implementation of further excitation sources, e.g., a 4  μm interband cascade laser, to target the strong mid-infrared absorption bands of CO2.28

Disclosures

The authors declare no conflict of interest. Relevant data can be made available by the author upon request.

Code and Data Availability

All data can be made available by the author upon request to manuel.tanzer@tugraz.at.

Acknowledgments

The work presented here was conducted in the framework of the FFG funded project “Green Sensing” and the NATO SPS program “Photonic Nano Particle Sensors for Detecting CBRN events.” The work was also supported by TU Graz Open Access Publishing Fund. Part of the content of this work has previously been presented at the SPIE Optics and Photonics conference in 2023 and can be found in the respective proceedings.29

References

1. 

D. Popa and F. Udrea, “Towards integrated mid-infrared gas sensors,” Sensors, 19 2076 https://doi.org/10.3390/s19092076 SNSRES 0746-9462 (2019). Google Scholar

2. 

A. Fathy et al., “Direct absorption and photoacoustic spectroscopy for gas sensing and analysis: a critical review,” Laser Photonics Rev., 16 2100556 https://doi.org/10.1002/lpor.202100556 (2022). Google Scholar

3. 

J. Hodgkinson and R. Tatam, “Optical gas sensing: a review,” Meas. Sci. Technol., 24 012004 https://doi.org/10.1088/0957-0233/24/1/012004 MSTCEP 0957-0233 (2012). Google Scholar

4. 

K. Krzempek, “A review of photothermal detection techniques for gas sensing applications,” Appl. Sci., 9 2826 https://doi.org/10.3390/app9142826 (2019). Google Scholar

5. 

A. J. Campillo et al., “Fabry-Perot photothermal trace detection,” Appl. Phys. Lett., 41 327 –329 https://doi.org/10.1063/1.93524 APPLAB 0003-6951 (1982). Google Scholar

6. 

G. Dudzik and K. Abramski, “Solid-state laser intra-cavity photothermal sensor (slips) for gas detection with parts-per-billion sensitivity,” Eng. Proc., 21 34 https://doi.org/10.3390/engproc2022021034 EASMD5 (2022). Google Scholar

7. 

H. Moosmüller and W. Arnott, “Folded Jamin interferometer: a stable instrument for refractive-index measurements,” Opt. Lett., 21 438 –440 https://doi.org/10.1364/OL.21.000438 OPLEDP 0146-9592 (1996). Google Scholar

8. 

M. Islam et al., “Chronology of Fabry-Perot interferometer fiber-optic sensors and their applications: a review,” Sensors, 14 7451 –7488 https://doi.org/10.3390/s140407451 SNSRES 0746-9462 (2014). Google Scholar

9. 

P. Huangfu and R. Atkinson, “Long-term exposure to NO2 and O3 and all-cause and respiratory mortality: a systematic review and meta-analysis,” Environ. Int., 144 105998 https://doi.org/10.1016/j.envint.2020.105998 ENVIDV 0160-4120 (2020). Google Scholar

10. 

W. H. Organization, “Who global air quality guidelines,” https://www.who.int/publications/i/item/9789240034228 (2021). Google Scholar

11. 

P. Breitegger, B. Lang and A. Bergmann, “Intensity modulated photothermal measurements of NO2 with a compact fiber-coupled Fabry–Pérot interferometer,” Sensors, 19 3341 https://doi.org/10.3390/s19153341 SNSRES 0746-9462 (2019). Google Scholar

12. 

J. P. Waclawek et al., “Balanced-detection interferometric cavity-assisted photothermal spectroscopy,” Opt. Express, 27 12183 –12195 https://doi.org/10.1364/OE.27.012183 OPEXFF 1094-4087 (2019). Google Scholar

13. 

P. Patimisco et al., “Recent advances in quartz enhanced photoacoustic sensing,” Appl. Phys. Rev., 5 011106 https://doi.org/10.1063/1.5013612 (2018). Google Scholar

14. 

G. Leahu et al., “Trace gas analysis from glazes by means of a compact photothermal deflection spectroscopy apparatus,” Rev. Sci. Instrum., 84 123111 https://doi.org/10.1063/1.4848875 RSINAK 0034-6748 (2013). Google Scholar

15. 

M. Bertolotti and R. L. Voti, “A note on the history of photoacoustic, thermal lensing, and photothermal deflection techniques,” J. Appl. Phys., 128 230901 https://doi.org/10.1063/5.0023836 JAPIAU 0021-8979 (2020). Google Scholar

16. 

M. Tanzer, B. Lang and A. Bergmann, “Fabrication of a low-cost, fiber-coupled, and air-spaced Fabry-Pérot etalon,” J. Vis. Exp., 192 e65174 https://doi.org/10.3791/65174 (2023). Google Scholar

17. 

J. Chen et al., “Acoustic performance study of fiber-optic acoustic sensors based on Fabry-Pérot etalons with different Q factors,” Micromachines, 13 118 https://doi.org/10.3390/mi13010118 (2022). Google Scholar

18. 

S. E. Bialkowski, N. G. C. Astrath and M. A. Proskurnin, Photothermal Spectroscopy Methods, 2nd ed.Wiley & Sons Ltd.( (2019). Google Scholar

19. 

P. Kluczynski et al., “Wavelength modulation absorption spectrometry – an extensive scrutiny of the generation of signals,” Spectrochim. Acta Part B: At. Spectrosc., 56 1277 –1354 https://doi.org/10.1016/S0584-8547(01)00248-8 (2001). Google Scholar

20. 

I. E. Gordon et al., “The HITRAN2020 molecular spectroscopic database,” J. Quantum Spectrosc. Radiat. Transf., 277 107949 https://doi.org/10.1016/j.jqsrt.2021.107949 (2022). Google Scholar

21. 

M. Vaughan, The Fabry-Perot Interferometer: History, Theory, Practice and Applications, 1st ed.Taylor & Francis Group( (1989). Google Scholar

22. 

R. Drever et al., “Laser phase and frequency stabilization using an optical resonator,” Appl. Phys. B, 31 97 –105 https://doi.org/10.1007/BF00702605 (1983). Google Scholar

23. 

P. Breitegger and A. Bergmann, “A precise gas dilutor based on binary weighted critical flows to create NO2 concentrations,” MDPI Proc., 19 998 https://doi.org/10.3390/proceedings2130998 (2018). Google Scholar

24. 

P. Werle, R. Mücke and F. Slemr, “The limits of signal averaging in atmospheric trace-gas monitoring by tunable diode-laser absorption spectroscopy (TDLAS),” Appl. Phys. B, 57 131 –139 https://doi.org/10.1007/BF00425997 (1993). Google Scholar

25. 

W. J. Riley, “Handbook of frequency stability analysis,” https://tsapps.nist.gov/publication/get_pdf.cfm?pub_id=50505 (2008). Google Scholar

26. 

M. Baudelet, Laser Spectroscopy for Sensing: Fundamentals, Techniques and Applications, Elsevier( (2014). Google Scholar

27. 

C. D. Motchenbacher and J. A. Connelly, Low-Noise Electronic System Design, Wiley & Sons Ltd. ( (1993). Google Scholar

28. 

Z. Du et al., “Mid-infrared tunable laser-based broadband fingerprint absorption spectroscopy for trace gas sensing: a review,” Appl. Sci., 9 338 https://doi.org/10.3390/app9020338 (2019). Google Scholar

29. 

M. Tanzer, B. Lang and A. Bergmann, “Photothermal gas sensing with a custom-made, fiber-coupled Fabry-Pérot etalon,” Proc. SPIE, 12672 126720D https://doi.org/10.1117/12.2676337 PSISDG 0277-786X (2023). Google Scholar

Biography

Manuel Tanzer studied technical physics and physical energy and measurement engineering at the Vienna University of Technology (Austria). Since 2020, he is a PhD candidate at the Institute of Electrical Measurement and Sensor Systems at the Graz University of Technology (Austria). His research topic is the miniaturization of photothermal gas and aerosol sensors. He is a student-member of SPIE.

Benjamin Lang is a university assistant at the Institute for Electrical Measurement and Sensor Systems, Graz University of Technology (Austria). He received his master’s degree in technical physics in 2015 and his PhD from the Faculty of Electrical and Information Engineering in 2020, both at Graz University of Technology. He is interested in and researching laser-based optical sensing techniques, focusing on laser absorption spectroscopy (tunable diode, photoacoustic, and photothermal) in gasses and condensed phases.

Matej Njegovec is an assistant professor at the University of Maribor (Slovenia) and received his BS degree in 2008. In 2013, he received his PhD in electrical engineering from the University of Maribor. His main research includes fiber optic sensors, optical sensor interrogation techniques, and the design of optoelectronic systems. He has excellent background in analog and RF circuit design, semiconductor light sources, optoelectronic components, modulation techniques, signal-processing of modulated light signals, and general signal integration techniques related to optical sensors. His biggest motive is to find innovative and cost-efficient solutions for fiber-optic sensors that would allow for the introduction of fiber-sensor technology into a broader range of practical applications.

Vedran Budinski received his BSc and PhD degrees from the Faculty of Electrical Engineering and Computer Science at the University of Maribor, Slovenia, in 2012 and 2017, respectively. Since 2012, he has been working as a researcher in the Laboratory for Electro-Optical and Sensor Systems at the University of Maribor. His research interests include fiber optic sensors for measuring twist/rotation, fiber optic microfluidic sensors, and design of miniature microstructures on the tip of an optical fiber, employing various micromachining techniques for detecting various physical and chemical parameters.

Simon Pevec is an assistant professor at the University of Maribor. In 2014, he received his PhD in electrical engineering from the University of Maribor. He has more than 15 years of researching experiences in field of photonics, optoelectronics, MOEMS, sensors, and measurements. His greatest experiences are related with fiber based Fabry–Perot sensors, multiparameter sensors, optoelectronics system design, and micromachining of optical fibers including selective etching; all to develop all kind of miniature all-fiber Fabry–Pérot sensors and sensors systems.

Jure Javornik finished his bachelor’s degree (2019) and master’s degree (2022) in electrical engineering at FERI UM. During the master’s degree, he was working on femtosecond laser machining procedures for fiber-optic sensor fabrication. Upon the degree completion, he started working as a junior researcher at Laboratory for Electro-Optical and Sensor Systems (LEOSS), stationed at FERI UM, where he is currently working on his PhD. His field of research includes femtosecond laser micromachining, selective etching, type 2 FBG inscription, fiber-optic sensor design, fabrication, and cost-efficient interrogation techniques.

Denis Donlagic is full professor at the University of Maribor, received his first PhD in 1998 at the University of Ljubljana (Slovenia). In 2000, he finished his second PhD at the University of Strathclyde (Glasgow, United Kingdom). He has a broad background in the fields of photonics systems, opto-electronics systems, optical fibers, and MOEMS. In particular, he has research and development experience in the fields of optical fiber sensors, optical fibers, and opto-electronics system design. His focus is in finding new scientific paradigms and technology concepts while bringing them into practical applications. He is member of IEEE and member of the editorial board of the Journal of Lightwave Technology.

Alexander Bergmann is professor for electronic sensor systems and head of the Institute of Electrical Measurement and Sensor Systems at Graz University of Technology. His research is dedicated to sensor effects, sensor materials, and sensor systems. One focus area is the quantification of fluid as well as aerosol properties (air polluters and climate forcers) with emphasis on high temporal resolution. Additionally, he is laboratory head of the Christian Doppler Laboratory for Structured Matter-based Sensing, where he and his team investigate highly sensitive sensing structures based on structured light and matter.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Manuel Tanzer, Benjamin Lang, Matej Njegovec, Vedran Budinski, Simon Pevec, Jure Javornik, Denis Donlagic, and Alexander Bergmann "Photothermal gas sensing with a custom-made, fiber-coupled Fabry–Pérot etalon," Optical Engineering 63(6), 064101 (4 June 2024). https://doi.org/10.1117/1.OE.63.6.064101
Received: 23 December 2023; Accepted: 20 May 2024; Published: 4 June 2024