We report a giant enhancement of Spin Hall (SH) shift even for normal incidence in an exotic optical system, an inhomogeneous anisotropic medium having complex spatially varying birefringent structure. The spatial variation of birefringence is obtained by changing the three dimensional orientation of liquid crystal by modulating the pixels with user-controlled greyscale value. This polarization dependent spatial variation (in a plane transverse to the direction of propagation of light) of the transmitted light beam (for incident fundamental Gaussian beam lacking any intrinsic angular momentum) through such inhomogeneous anisotropic medium was recorded using an Eigenvalue calibrated Stokes- Mueller imaging system. Giant SH shift was manifested as distinctly different spatial distribution of the recorded output Stokes vector elements for two orthogonal (left and right) input circular polarization states. We unravel the reason for such large enhancement of SH shift by performing rigorous three dimensional analysis of polarization evolution in such complex anisotropic medium. The theoretical analysis revealed that generation of large magnitude of transverse energy flow (quantified via the Poynting vector evolution inside the medium) originating from Spin Orbit Interaction (SOI) in the inhomogeneous birefringent medium leads to the observation of such a large spin dependent deflection of the trajectory of light beam.
We present a novel spectral Mueller matrix measurement system for both elastic and inelastic scattering (fluorescence)
polarimetric measurements. The system comprises of a Xenon lamp as excitation source, a polarization state generator
(PSG) and a polarization state analyzer (PSA) unit to generate and analyze polarization states required for 4 x 4 sample
Mueller matrix measurements, coupled to a spectrometer for spectrally resolved (λ ~ 400 - 800 nm) signal detection.
The PSG unit comprises of a fixed linear polarizer (polarization axis oriented at horizontal position) followed by a
rotatable broadband quarter wave plate. The sample-scattered light is collected and collimated using an assembly of
lenses, then passes through the PSA unit, and is finally recorded using the spectrometer. The PSA unit essentially
consists of a similar arrangement as that of the PSG, but positioned in reverse order, and with the axis of the linear
polarizer oriented at vertical position. A sequence of sixteen measurements are performed by changing the orientation of
the fast axis of the quarter wave plates of the PSG unit (for generating the four required elliptical polarization states) and
that of the PSA unit (for analyzing the corresponding polarization states). The orientation angles
(35°, 70°, 105° and 140°) were chosen based on optimization of the PSG and PSA matrices to yield most stable system
Mueller matrices. The performance of the polarimeter was calibrated using Eigenvalue calibration method which also
yielded the actual values of the system PSG and PSA matrices at each wavelength. The system has been automated and
is capable of Mueller matrix measurement with high accuracy over the entire spectral range 400 - 800 nm (elemental
error < 0.01). For recording the elastic scattering Mueller matrix of sample, the PSG and PSA matrices for each
wavelength are used, while for fluorescence Mueller matrix measurements, the PSG for the excitation wavelength
(chosen to be 405 nm) and PSA for varying emission wavelengths (450 - 800 nm) are used. The developed spectral
Mueller matrix system has been initially used to record both elastic scattering and fluorescence Mueller matrices from
normal and cancerous cervical tissues.
Access to the requested content is limited to institutions that have purchased or subscribe to SPIE eBooks.
You are receiving this notice because your organization may not have SPIE eBooks access.*
*Shibboleth/Open Athens users─please
sign in
to access your institution's subscriptions.
To obtain this item, you may purchase the complete book in print or electronic format on
SPIE.org.
INSTITUTIONAL Select your institution to access the SPIE Digital Library.
PERSONAL Sign in with your SPIE account to access your personal subscriptions or to use specific features such as save to my library, sign up for alerts, save searches, etc.