Marcin Masalski, Adam Wąż, Przemysław Błauciak, Tomasz Zatonski, Krzysztof Morawski
Journal of Biomedical Optics, Vol. 26, Issue 07, 077001, (July 2021) https://doi.org/10.1117/1.JBO.26.7.077001
TOPICS: Laser beam diagnostics, Bone, Vibrometry, Ear, Optical fibers, Laser Doppler velocimetry, Surgery, Reflectivity, Fiber lasers, In vivo imaging
Significance: Measurements of auditory ossicles displacement are commonly carried out by means of laser-Doppler vibrometry (LDV), which is considered to be a gold standard. The limitation of the LDV method, especially for in vivo measurements, is the necessity to expose an object in a straight line to a laser beam operating from a distance. An alternative to this approach is the use of a handheld laser-fiber vibrometry probe (HLFVP) with a curved tip.
Aim: We evaluate the feasibility of an HLFVP with a curved tip for measuring sound-induced displacement of the auditory ossicles.
Approach: A handheld vibrometer probe guiding the laser beam with a fiber-optic cable was used for displacement measurements of the incus body and the posterior crus of the stapes. Tonal stimuli at frequencies of 0.5, 1, 2, and 4 kHz were presented by means of an insert earphone positioned in the outer ear canal. The probe was fixed at the measurement site using a tripod or hand-held by one of the two surgeons.
Results: The measurements were carried out on six fresh temporal bones. Multivariate analysis of variance showed statistically significant differences for stimulus frequency (F3,143 = 29.37, p < 0.001, and η2 = 0.35), bone (F5,143 = 4.61, p = 0.001, and η2 = 0.01), and measurement site (F1,143 = 4.74, p = 0.03, and η2 = 0.02) in the absence of statistically significant differences for the probe fixation method (F2,143 = 0.15, p = 0.862, and η2 = 0.001). Standard deviations of the means were 6.9, 2.6, 1.9, and 0.6 nm / Pa for frequency, bone, site, and fixation, respectively. Ear transfer functions were found to be consistent with literature data.
Conclusions: The feasibility of applying HLFVP to measure the displacement of auditory ossicles has been confirmed. HLFVP offers the possibility of carrying out measurements at various angles; however, this needs to be standardized taking into account anatomical limitations and surgical convenience.